SENSIT s.r.o. is one of the most important manufacturers of temperature sensors in the temperature range from -190°C to 600°C on the Czech market.

Its products are used in the field of metering and control, in the field of billing measurement of heat consumption and in all sectors of industry, agriculture and services where it is necessary to monitor temperature as an important parameter of technological processes.

Temperature sensors with head up to 600°C

Description and usage
These resistance temperature sensors are designed for contact temperature measurement of gaseous substances in outdoor or industrial areas. The plastic head is equipped with a cable gland (a terminal block is located in the head) or a connector. The standard temperature range of the sensors is - 50 to 100 ° C. The sensors can be used for all control systems that are compatible with the types of sensors or output signals listed in the table of technical parameters. Easy installation of temperature sensors is ensured thanks to the unique " S head " design of SENSIT s.r.o. Temperature sensors are designed for operation in a non-aggressive environment. ACCESSORIES DECLARATION, CERTIFICATION, CALIBRATION
  • The manufacturer issues the EU Declaration of Conformity.
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
Note: Temperature sensors with a 4-20 mA output signal and frequency output are only available in variant A.
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached files
NS 110x NS 110xK NS 111x NS 111xK NS 112x NS 112xK NS 310x NS 310xK NS 311x NS 311xK 1 mA 1 mA 1 mA 0.3 mA 0.3 mA
Sensor type (K - with connector)
Sensor type Ni 1000/5000 Ni 1000/6180 Ni 891 Ni 10000/5000 Ni 10000/6180
Measuring range - 50 to 100 ° C
Max. dc measuring current
NS 113x NS 113xK HS 110x HS 110xK 0.7 mA 1.5 mA 1 mA
Sensor type (K - with connector) PTS 110x PTS 110xK PTS 210x PTS 210xK PTS 310x PTS 310xK
Sensor type T1 = Ni 2226 PT 100/3850 PT 500/3850 PT 1000/3850 NTC thermistor 20 kΩ
Measuring range - 50 to 100 ° C
Max. dc measuring current 3 mA 1 mW *)
*) maximum power consumption NS 510A NS 510AK NS 710x NS 710xK Pt 1000/3850
Sensor type (K - with connector) Note
Sensor type PT1000 / 3850
Output signal 4 to 20 mA 0 to 10 V
Standard measuring ranges **) - 30 to 60 ° C 0 to 35 ° C 0 to 100 ° C 0 to 150 ° C - 30 to 60 ° C 0 to 35 ° C 0 to 100 ° C 0 to 150 ° C Temperature around the head -30 to 70 ° C; ambient temperature sensor head NS 810A (K) -30-70 ° C
Supply voltage (U nap ) 11 to 30 V DC 15 to 30 V DC recommended value 24 V DC, recommended power supply 12 V DC for NS 810A (K) Axima AXSP3PO2012
Load resistance Rz 150 Ω for Unap = 12 V 700 Ω for Unap = 24 V > 10kΩ
Sensor interruption output signal > 24 mA > 10.5 mA
Sensor short circuit output signal <3.5 mA ~ 0 V
Note: x = variant A or variant B **) Based on the customer's request, it is possible to provide a custom measuring range in the range from -40 to 150 ° C, while the minimum range of the range must be 35 ° C (eg -20 to 15 ° C; -30 to 80 ° C, etc. )
OTHER PARAMETERS
Accuracy class Ni sensors: cl. B, t = ± (0.4 + 0.007t), for t ≥ 0; t = ± (0.4 + 0.028 | t |), for t ≤ 0 in ° C; Pt sensors: cl. B according to ČSN EN 60751, t = ± (0.3 + 0.005 | t |) in ° C NTC 20 kΩ: ± 1 ° C for the range 0 to 70 ° C
Measurement error Ns 510a (K), Ns 710x (K) and Ns 810a (K) NS 510 A ± 1 ° C depending on air flow rate NS 710 A 0.6% of range, min. 0.5 ° C NS 810A (K) 0.5 ° C for spans <100 ° C, 100 ° C
Sensor connection according to the wiring diagram
Stem length (variant A) resistive and voltage output 0 to 10 V: 25 mm current output 4 to 20 mA: 50 mm
Response time τ0.5 <9 s (in flowing air 1m s -1 ) - variant A τ0.5 ≤ 30 s (in flowing air 1m s -1 ) - variant B
Recommended wire cross-section - bushing sensors 0.35 to 1.5 mm 2
Head connector type - sensors with connector RSFM4 - Lumberg
Insulation resistance > 200 MΩ at 500 V DC, 25 ° ± 3 ° C; humidity <85%
Degree of coverage IP 65 according to ČSN EN 60529
Stem material stainless steel DIN 1.4301 - variant A
Head material POLYAMIDE
Working conditions ambient temperature: -50 to 100 ° C; -30 to 70 ° C with transducer; -30 to 70 ° C with freq. output
relative humidity: max. 100% (at an ambient temperature of 25 ° C)
atmospheric pressure: 70 to 107 kPa
Weight about 0.15 kg
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
These resistance temperature sensors are designed for contact temperature measurement of gaseous substances in outdoor or industrial areas. The standard temperature range of the sensors is -30 ° C to 100 ° C. The sensors can be used for all control systems that are compatible with the types of sensors or output signals listed in the table of technical parameters. Temperature sensors are designed for operation in a non-aggressive environment. ACCESSORIES DECLARATION, CERTIFICATION, CALIBRATION
  • The manufacturer issues an EU Declaration of Conformity.
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached files
NK 110 NK 111 NK 112 NK 310 NK 311 Ni 891 Ni 1 mA 1 mA 1 mA 0.3 mA 0.3 mA
Sensor type
Sensor type Ni 1000/5000 Ni 1000/6180 Ni 10000/5000 Ni 10000/6180
Measuring range - 30 to 100 ° C
Max. dc measuring current
NK 113 HK 110 0.7 mA 1.5 mA 1 mA
Sensor type PTK 110 PTK 210 PTK 310
Sensor type T1 = Ni 226 PT 100/3850 PT 500/3850 PT 1000/3850 NTC thermistor 20 kΩ
Measuring range - 30 to 100 ° C
Max. dc measuring current 3 mA 1 mW *)
*) maximum power consumption NK 510 Pt 1000/3850
Sensor type Note
Sensor type
Output signal 4 to 20mA
Standard measuring ranges - 30 to 60 ° C 0 to 35 ° C 0 to 100 ° C 0 to 150 ° C Temperature around the head -30 to 70 ° C
Measurement error ± 1 ° C depending on air velocity
Supply voltage (Unap) 11 to 30 V DC recommended value 24 V DC
Load resistance Rz 150 Ω for Unap = 12 V 700 Ω for Unap = 24 V
Sensor interruption output signal > 24 mA
Sensor short circuit output signal <3.5 mA
OTHER PARAMETERS
Accuracy class Ni sensors: cl. B, t = ± (0.4 + 0.007t), for t ≥ 0; t = ± (0.4 + 0.028 | t |), for t ≤ 0 in ° C; Pt sensors: cl. B according to ČSN EN 60751, t = ± (0.3 + 0.005 | t |) in ° C NTC 20 kΩ: ± 1 ° C for the range 0 to 70 ° C
Sensor connection according to the wiring diagram
Stem length L resistive output: 25 mm current output 4 to 20 mA: 50 mm
Response time τ0.5 <9 s (in flowing air 1m s -1 )
Recommended wire cross section 0.35 to 1.5 mm 2
Insulation resistance > 200 MΩ at 500 V DC, 25 ° ± 3 ° C; humidity <85%
Degree of coverage IP 54 according to ČSN EN60529
Stem material stainless steel DIN 1.4301
Head type LIMATHERM MA
Head material aluminum alloy
Working conditions ambient temperature: -30 to 100 ° C; -30 to 70 ° C with transducer
relative humidity: max. 100% (at an ambient temperature of 25 ° C)
atmospheric pressure: 70 to 107 kPa
Weight about 0.2 kg
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
These resistance temperature sensors are designed for contact temperature measurement of liquid or gaseous substances. The plastic head is equipped with a cable gland (a terminal block is located in the head) or a connector. In combination with the central holder, the sensor is suitable for measuring the temperature in air conditioning ducts. The combination of sensor and well is suitable for measuring in pipes. A modified variant of the sensor with welded-on screw connection is suitable for direct measurement of the medium in the pipeline. The standard temperature range for using temperature sensors is -30 ° C to 150 ° C. By extending the length of the stem, it is possible to use temperature sensors up to temperatures of 250 ° C. Temperature sensors can be used for all control systems that are compatible with the types of sensors or output signals listed in the table of technical parameters. Easy installation of temperature sensors is ensured thanks to the unique " S head " design of SENSIT s.r.o. The sensors are designed for operation in a non-aggressive environment. ACCESSORIES DECLARATION, CERTIFICATION, CALIBRATION
  • The manufacturer issues an EU Declaration of Conformity.
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached files
NS 120 NS 120K NS 121 NS 121K NS 122 NS 122K NS 320 NS 320K NS 321 NS 321K 1 mA 1 mA 1 mA 0.3 mA 0.3 mA
Sensor type (K - with connector)
Sensor type Ni 1000/5000 Ni 1000/6180 Ni 891 Ni 10000/5000 Ni 10000/6180
Measuring range -30 to 150 ° C
Max. dc measuring current
NS 123 NS 123K HS 120 HS 120K Pt 500/3850 Pt 1000/3850 0.7 mA 1.5 mA 1 mA
Sensor type (K - with connector) PTS 120 PTS 120K PTS 220 PTS 220K PTS 320 PTS 320K
Sensor type T1 = Ni 2226 Pt 100/3850 NTC thermistor 20 kΩ
Measuring range -30 to 150 ° C - 50 to 150 ° C (temperature around the head - 30 to 100 ° C) -30 to 150 ° C
Max. dc measuring current 3 mA 10 mW *)
*) maximum power consumption NS 520 NS 520K NS 720 NS 720K Pt 1000/3850 Pt 1000/3850 -td to 50 ° C -30 to 60 ° C 0 to 35 ° C 0 to 100 ° C 0 to 150 ° C 0 to 200 ° C 0up to 250 ° C -td to 60 ° C 0 to 35 ° C 0 to 100 ° C 0 to 150 ° C 0 to 200 ° C 0 to 250 ° C
Sensor type (K - with connector) Note
Sensor type
Output signal 4 to 20 mA 0 to 10 V
Standard measuring ranges **) temperature around the head -30 to 70 ° C;
Supply voltage (Unap) 11 to 30 V DC 15 to 30 V DC recommended value 24 V DC; recommended power supply 12 V DC NS 840 (K) Axima AXSP3P02012
Load resistance Rz 150 Ω for UNAP = 12 V 700 Ω for UNAP = 24 V > 10 kΩ
Sensor interruption output signal > 24 mA > 10.5 V
Sensor short circuit output signal <3.5 mA ~ 0 V
**) Based on the customer's request, it is possible to provide a custom measuring range in the range from -40 to 150 ° C, while the minimum range of the range must be 35 ° C (eg -20 to 15 ° C; -30 to 80 ° C, etc. )
OTHER PARAMETERS
Sensor accuracy class Ni sensors: cl. B, Δt = ± (0.4 + 0.007t), for t ≥ 0; Δt = ± (0.4 + 0.028 | t |), for t ≤ 0 in ° C; Pt sensors: cl. B according to ČSN EN 60751, Δt = ± (0.3 + 0.005 | t |) in ° C NTC 20 kΩ: ± 1 ° C for the range 0 to 70 ° C
NS 520 (K), NS 720 (K), and NS 820 (K) measurement error <0.6% of range, minimum 0.5 ° C NS 820 (K) 0.5 ° C for range <100 ° C, <0.6% of range for range <100 ° C
Sensor connection according to the scheme
Standard L1 stem length 70, 120, 180, 240, 300, 360, 420 mm
Response time τ0.5 <9 s (in flowing water 0.4 m.s -1 )
Recommended wire cross-section - bushing sensors 0.35 to 1.5 mm 2
Head connector type - sensors with connector RSFM4 - Lumberg
Insulation resistance > 200 MΩ at 500 V DC, 25 ° ± 3 ° C; humidity <85%
Degree of coverage IP 65 according to ČSN EN 60529
Stem material stainless steel DIN 1.4301
Head material POLYAMIDE
Working conditions ambient temperature: -30 to 100 ° C; -30 to 70 ° C with transducer; -30 to 70 ° C with freq. output relative humidity: max. 100% (at ambient temperature 25 ° C) atmospheric pressure: 70 to 107 kPa
Weight about 0.15 kg
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
These resistance temperature sensors are designed for contact temperature measurement of liquid or gaseous substances. In combination with the central holder, the temperature sensor is suitable for measuring the temperature in air conditioning ducts. The combination of temperature sensor and well is suitable for measuring in pipes. A modified variant of the temperature sensor with welded-on screw connection is suitable for direct measurement of the medium in the pipeline. The standard temperature range for measuring the use of sensors is -30 ° C to 200 ° C. By extending the length of the stem, it is possible to use temperature sensors up to temperatures of 250 ° C. Temperature sensors can be used for all control systems that are compatible with the types of sensors or output signals listed in the table of technical parameters. Temperature sensors are designed for operation in a non-aggressive environment. ACCESSORIES DECLARATION, CERTIFICATION, CALIBRATION
  • The manufacturer issues an EU Declaration of Conformity.
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached files
NK 120 NK 121 NK 122 NK 320 NK 321 1 mA 1 mA 1 mA 0.3 mA 0.3 mA
Sensor type
Sensor type Ni 1000/5000 Ni 1000/6180 Ni 891 Ni 10000/5000 Ni 10000/6180
Measuring range - 30 to 200 ° C (temperature around the head)
Max. dc measuring current
NK 123 HK 120 Pt 500/3850 Pt 1000/3850 0.7 mA 1.5 mA 1 mA
Sensor type PTK 120 PTK 220 PTK 320
Sensor type T1 = Ni 2226 Pt 100/3850 NTC thermistor 20 kΩ
Measuring range -30 to 150 ° C - 50 to 200 ° C (temperature around the head - 30 to 100 ° C) -30 to 150 ° C
Max. dc measuring current 3 mA 10 mW *)
*) maximum power consumption NK 520 Pt 1000/3850 -td to 50 ° C -30 to 60 ° C 0 to 35 ° C 0 to 100 ° C 0 to 150 ° C 0 to 200 ° C 0 to 250 ° C
Sensor type Note
Sensor type
Output signal 4 to 20 mA
Standard measuring ranges temperature around the head -30 to 70 ° C;
Measurement error <0.6% of range at least 0,5 ° C
Supply voltage (Unap) 11 to 30 V DC recommended value 24 V DC
Load resistance Rz 150 Ω for Unap = 12 V 700 Ω for Unap = 24 V
Sensor interruption output signal > 24 mA
Sensor short circuit output signal <3.5 mA
OTHER PARAMETERS
Sensor accuracy class Ni sensors: cl. B, Δt = ± (0.4 + 0.007t), for t ≥ 0; Δt = ± (0.4 + 0.028 | t |), for t ≤ 0 in ° C; Pt sensors:cl. B according to ČSN EN 60751, Δt = ± (0.3 + 0.005 | t |) in ° C NTC 20 kΩ: ± 1 ° C for the range 0 to 70 ° C
Sensor connection according to the wiring diagram
Standard stem length 70, 120, 180, 240, 300, 360, 420 mm
Response time τ0.5 <9 s (in flowing water 0.4 m.s -1 )
Recommended wire cross section 0.35 to 1.5 mm 2
Insulation resistance > 200 MΩ at 500 V DC, 25 ° ± 3 ° C; humidity <85%
Degree of coverage IP 54 according to ČSN EN 60529
Stem material stainless steel DIN 1.4301
Head type LIMATHERM MA
Head material aluminum alloy
Working conditions ambient temperature: -30 to 100 ° C; -30 to 70 ° C with transducer relative humidity: max. 100% (at ambient temperature 25 ° C) atmospheric pressure: 70 to 107 kPa
Weight about 0.15 kg
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
These resistance temperature sensors are designed for contact temperature measurement of liquid or gaseous substances. The plastic head is equipped with a cable gland (a terminal block is located in the head) or a connector. The design of the sensor stem allows the use of the sensor for direct measurement of temperature in pipes and at the same time ensures a fast response of the sensor to temperature changes. The standard temperature range of the sensors is -30 to 130 ° C. The sensors can be used for all control systems that are compatible with the types of sensors or output signals listed in the table of technical parameters. Easy installation of temperature sensors is ensured thanks to the unique " S head " design of SENSIT s.r.o. Temperature sensors are designed for operation in a non-aggressive environment. ACCESSORIES DECLARATION, CERTIFICATION, CALIBRATION
  • The manufacturer issues an EU Declaration of Conformity.
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached files
NS 160 NS 160K NS 161 NS 161K NS 162 NS 162K 1 mA 1 mA 1 mA
Sensor type (K - with connector)
Sensor type Ni 1000/5000 Ni 1000/6180 Ni 891
Measuring range - 30 to 130 ° C (temperature around the head -30 to 100 ° C)
Max. dc measuring current
PTS 360 PTS 360K HS 160 HS 160K Pt 500/3850 Pt 1000/3850 1.5 mA 1 mA
Sensor type (K - with connector) PTS 160 PTS 160K PTS 260 PTS 260K
Sensor type Pt 100/3850 NTC thermistor 20 kΩ
Measuring range - 50 to 130 ° C (temperature around the head - 30 to 100 ° C) -30 to 130 ° C
Max. dc measuring current 3 mA 10 mW *)
*) maximum power consumption NS 560 NS 560K NS 760 NS 760K Pt 1000/3850 Pt 1000/3850 -td to 50 ° C -30 to 60 ° C 0 to 35 ° C 0 to 100 ° C 0 to 150 ° C -td to 60 ° C 0 to 35 ° C 0 to 100 ° C 0 to 150 ° C
Sensor type (K - with connector) Note
Sensor type
Output signal 4 to 20 mA 0 to 10 V
Standard measuring ranges **) temperature around the head -30 to 70 ° C; temperature around the sensor head NS 820 (K) -30 to 70 ° C
Supply voltage (Unap) 11 to 30 V DC 15 to 30 V DC recommended value 24 V DC; recommended power supply 12 V DC NS 820 (K) Axima AXSP3P02012
Load resistance Rz 150 Ω for Unap = 12 V 700 Ω for Unap = 24 V > 10 kΩ
Sensor interruption output signal > 24 mA > 10.5 V
Sensor short circuit output signal <3.5 mA ~ 0 V
**) Based on the customer's request, it is possible to provide a custom measuring range in the range from -40 to 150 ° C, while the minimum range of the range must be 35 ° C (eg -20 to 15 ° C; -30 to 80 ° C andunder.) NS 560 and NS 760 <0.6% of range, minimum 0.5 ° C NS 860 (K) 0.5 ° C for range <100 ° C, 100 ° C 4 ± 0.1 mm
OTHER PARAMETERS
Sensor accuracy class Ni sensors: cl. B, Δt = ± (0.4 + 0.007t), for t ≥ 0; Δt = ± (0.4 + 0.028 | t |), for t ≤ 0 in ° C; Pt sensors: cl. B according to ČSN EN 60751, Δt = ± (0.3 + 0.005 | t |) in ° C NTC 20 kΩ: ± 1 ° C for the range 0 to 70 ° C
Measurement error NS 560, NS 760, NS 860 (K)
Sensor connection according to the wiring diagram
Standard L1 stem length 50, 100, 160, 220 mm
Sensor Stem Diameter
Standard thread G 1/2 ″
Nominal sensor stem pressure PN 25
Response time τ0.5 <4 s (in flowing water 0.4 m.s -1 )
Recommended wire cross-section - bushing sensors 0.35 to 1.5 mm 2
Head connector type - sensors with connector RSFM4 - Lumberg, M12
Insulation resistance > 200 MΩ at 500 V DC, 25 ° ± 3 ° C; humidity <85%
Degree of coverage IP 65 according to ČSN EN 60529
Stem material stainless steel DIN 1.4301
Head material POLYAMIDE
Working conditions ambient temperature: -30 to 100 ° C; -30 to 70 ° C with transducer relative humidity: max. 100% (at ambient temperature 25 ° C) atmospheric pressure: 70 to 107 kPa
Weight about 0.15 kg
6 / 0.6 3.2 / 0.4 1.0 / 0.25
MAXIMUM FLOW RATE OF THE MEASURED MEDIA - AIR AND WATER STEAM / WATER [m.s -1 ]
Stem length up to 60 mm > 60 to 100 mm > 100 to 160 mm > 160 to 220 mm
Values ​​for 4 mm stem diameter 8 / 0.8
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
These resistance temperature sensors are designed for contact temperature measurement of liquid or gaseous substances. The design of the sensor stem allows the use of the sensor for direct measurement of temperature in pipes and at the same time ensures a fast response of the sensor to temperature changes. The standard temperature range of the sensors is -30 to 130 ° C. The sensors can be used for all control systems that are compatible with the types of sensors or output signals listed in the table of technical parameters. The sensors are designed for operation in a non-aggressive environment. DECLARATION, CERTIFICATION, CALIBRATION
  • The manufacturer issues an EU Declaration of Conformity.
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached files
NK 160 NK 161 NK 162 1 mA 1 mA 1 mA
Sensor type
Sensor type Ni 1000/5000 Ni 1000/6180 Ni 891
Measuring range - 30 to 130 ° C (temperature around the head -30 to 100 ° C)
Max. dc measuring current
HK 160 Pt 500/3850 Pt 1000/3850 1.5 mA 1 mA
Sensor type PTK 160 PTK 260 PTK 360
Sensor type Pt 100/3850 NTC thermistor 20 kΩ
Measuring range - 50 to 130 ° C (temperature around the head - 30 to 100 ° C) -30 to 130 ° C
Max. dc measuring current 3 mA 10 mW *)
*) maximum power consumption NK 560 Pt 1000/3850 -td to 50 ° C -30 to 60 ° C 0 to 35 ° C 0 to 100 ° C 0 to 150 ° C
Sensor type Note
Sensor type
Output signal 4 to 20 mA
Standard measuring ranges temperature around the head -30 to 70 ° C
Supply voltage (Unap) 11 to 30 V DC recommended value 24 V DC
Load resistance Rz 150 Ω for Unap = 12 V 700 Ω for Unap = 24 V
Sensor interruption output signal > 24 mA
Sensor short circuit output signal <3.5 mA
4 ± 0.1 mm
OTHER PARAMETERS
Sensor accuracy class Ni sensors: cl. B, Δt = ± (0.4 + 0.007t), for t ≥ 0; Δt = ± (0.4 + 0.028 | t |), for t ≤ 0 in ° C; Pt sensors: cl. B according to ČSN EN 60751, Δt = ± (0.3 + 0.005 | t |) in ° C NTC 20 kΩ: ± 1 ° C for the range 0 to 70 ° C
NS 560 measurement error <0.6% of range, minimum 0.5 ° C
Sensor connection according to the wiring diagram
Standard L1 stem length 50, 100, 160, 220 mm
Sensor Stem Diameter
Standard thread G 1/2 ″
Nominal sensor stem pressure PN 25
Response time τ0.5 <4 s (in flowing water 0.4 m.s -1 )
Recommended wire cross section 0.35 to 1.5 mm 2
Insulation resistance > 200 MΩ at 500 V DC, 25 ° ± 3 ° C; humidity <85%
Degree of coverage IP 54 according to ČSN EN 60529
Stem material stainless steel DIN 1.4301
Head material and type aluminum alloy, LIMATHERM MA
Working conditions ambient temperature: -30 to 100 ° C; -30 to 70 ° C with transducer relative humidity: max. 100% (at ambient temperature 25 ° C) atmospheric pressure: 70 to 107 kPa
Weight about 0.2 kg
6 / 0.6 3.2 / 0.4 1.0 / 0.25
MAXIMUM FLOW RATE OF THE MEASURED MEDIA - AIR AND WATER STEAM / WATER [m.s -1 ]
Stem length up to 60 mm > 60 to 100 mm > 100 to 160 mm > 160 to 220 mm
Values ​​for 4 mm stem diameter 8 / 0.8
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
MINI series resistance sensors are designed for measuring the temperature of liquid or gaseous substances in the temperature range -30 to 150 ° C. The sensors are equipped with a cable gland (a terminal block is located in the head) or a connector. In combination with the central holder, the sensor is suitable for measuring the temperature in air conditioning ducts. The combination of sensor and well is suitable for measuring in pipes. A modified variant of the sensor with welded-on screw connection is suitable for direct measurement of the medium in the pipeline. By extending the length of the stem, it is possible to use sensors up to temperatures of 250 ° C. The sensors can be used for all control systems that are compatible with the types of sensors or output signals listed in the table of technical parameters. Temperature sensors are designed for operation in a non-aggressive environment. ACCESSORIES DECLARATION, CERTIFICATION, CALIBRATION
  • The manufacturer issues an EU Declaration of Conformity.
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached files
MINI N 120 MINI N 120K MINI N 121 MINI N 121K MINI N 122 MINI N 122K MINI N 321 MINI N 321K 1 mA 1 mA 1 mA 0.3 mA 0.3 mA
Sensor type (K - with connector) MINI N 320 MINI N 320K
Sensor type Ni 1000/5000 Ni 1000/6180 Ni 891 Ni 10000/5000 Ni 10000/6180
Measuring range - 30 to 150 ° C (temperature around the head -30 to 100 ° C)
Max. dc measuring current
MINI N 123 MINI N 123K MINI P 120 MINI P 120K MINI P 220 MINI P 220K Pt 500/3850 0.7 mA 1.5 mA 1 mA
Sensor type (K - with connector) MINI P 320 MINI P 320K MINI H 120 MINI H 120K
Sensor type T1 = Ni 2226 Pt 100/3850 pT 1000/3850 NTC thermistor 20 kΩ
Measuring range -30 ° C to 150 ° C - 50 ° C to 150 ° C (temperature around the head -30 ° C to 100 ° C) -30 ° C to 150 ° C
Max. dc measuring current 3 mA 10 mW *)
*) maximum power consumption MINI N 520 Pt 1000/3850 -td to 50 ° C -30 to 60 ° C 0 to 35 ° C 0 to 100 ° C 0 to 150 ° C
Sensor type Note
Sensor type
Output signal 4 to 20 mA
Standard measuring ranges temperature around the head -30 to 70 ° C
Supply voltage (Unap) 11 to 30 V DC recommended value 24 V DC
Load resistance Rz 150 Ω for Unap = 12 V 700 Ω proUnap = 24 V
Sensor interruption output signal > 24 mA
Sensor short circuit output signal <3 mA
RSFM4 - Lumberg M12
OTHER PARAMETERS
Accuracy class Ni sensors: cl. B, Δt = ± (0.4 + 0.007t), for t ≥ 0; Δt = ± (0.4 + 0.028 | t |), for t ≤ 0 in ° C; Pt sensors: cl. B according to ČSN EN 60751, Δt = ± (0.3 + 0.005 | t |) in ° C NTC 20 kΩ: ± 1 ° C for the range 0 to 70 ° C
MINI N 520 measurement error <0.6% of range, minimum 0.5 ° C
Sensor connection according to the wiring diagram
Standard L1 stem length 70, 120, 180, 240, 300, 360, 420 mm
Response time τ0.5 <9 s (in flowing water 0.4 m.s -1 )
Terminal block type - bushing sensors Weco 951-A-LFDS, maximum conductor cross-section 1.5 mm 2
Connector Type - Connector Sensors
Power cord type - 4 to 20 mA transducers 2 x 0.25 mm 2 , PVC shielded, up to 80 ° C
Insulation resistance > 200 MΩ at 500 V DC, 25 ° ± 3 ° C; humidity <85%
Degree of coverage IP 65 according to ČSN EN 60529
Stem material stainless steel DIN 1.4301
Head material POLYAMIDE
Working conditions ambient temperature: -30 to 100 ° C; -30 to 70 ° C with converter or PVC cable relative humidity: max. 100% (at ambient temperature 25 ° C) atmospheric pressure: 70 to 107 kPa
Weight MINI approx. 60 g, MINI K 35 g
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
These resistance temperature sensors are designed for contact temperature measurement of liquid or gaseous substances. The sensors are manufactured in two variants: a small head with a resistance output signal and a large head with an output signal of 4 to 20 mA, as well as in versions with a bushing or with a connector. These sensors are primarily intended for use in the food industry. In combination with the central holder, the sensor is suitable for measuring the temperature in air conditioning ducts. The combination of sensor and well is suitable for measuring in pipes. The standard temperature range for the use of active sensors is -30 to 150 ° C, passive -30 to 200 ° C. The sensors can be used for all control systems that are compatible with the types of sensors or output signals listed in the table of technical parameters. Temperature sensors are designed for operation in a non-aggressive environment. ACCESSORIES DECLARATION, CERTIFICATION, CALIBRATION
  • The manufacturer issues an EU Declaration of Conformity.
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached files
NS 180P NS 180K NS 181P NS 181K NS 182P NS 182K NS 380P NS 380K NS 381P NS 381K 1 mA 1 mA 1 mA 0.3 mA 0.3 mA
Sensor type (K - with connector)
Sensor type Ni 1000/5000 Ni 1000/6180 Ni 891 Ni 10000/5000 Ni 10000/6180
Measuring range - 50 to 200 ° C (temperature around the head -30 to 100 ° C)
Max. dc measuring current
NS 183P NS 183K PTS 280P PTS 280K PTS 380P PTS 380K HS 180P HS 180K Pt 500/3850 0.7 mA 1.5 mA 1 mA
Sensor type (K - with connector) PTS 180P PTS 180K
Sensor type T1 = Ni 2226 Pt 100/3850 pT 1000/3850 NTC thermistor 20 kΩ
Measuring range -50 to 150 ° C - 50 to 200 ° C (temperature around the head -30 to 100 ° C) -30 to 150 ° C
Max. dc measuring current 3 mA 1 mW *)
*) maximum power consumption PTS 580P PTS 580K Pt 1000/3850 -td to 50 ° C -30 to 60 ° C 0 to 35 ° C 0 to 100 ° C 0 to 150 ° C
Sensor type (K - with connector) Note
Sensor type
Output signal 4 to 20 mA
Standard measuring ranges temperature around the head -30 to 70 ° C
Supply voltage (Unap) 11 to 30 V DC recommended value 24 V DC;
Load resistance Rz 150 Ω for Unap = 12 V 700 Ω for Unap = 24 V
Sensor signal output signal > > 24 mA
Sensor short circuit output signal <3.5 mA
OTHER PARAMETERS
Accuracy class Ni sensors: cl. B, Δt = ± (0.4 + 0.007t), for t ≥ 0; Δt = ± (0.4 + 0.028 | t |), for t ≤ 0 in ° C; Pt sensors: cl. B according to ČSN EN 60751, Δt = ± (0.3 + 0.005 | t |) in ° C NTC 20 kΩ: ± 1 ° C for the range 0 to 70 ° C
Measurement error <0.6% of range, minimum 0.5 ° C
Sensor connection according to the wiring diagram
Standard L1 stem length 70, 120, 180, 240 mm
Response time τ0.5 <9 s (in flowing water 0.4 m.s -1 )
Power cord - cable variant PVC shielded 2 x 0.25 mm 2
Head connector type - sensors with connector Lumberg RSFM4, M 12
Standard cable lengths 1, 2, 5, 10 m
Insulation resistance > 200 MΩ at 500 V DC, 25 ° ± 3 ° C; humidity <85%
Degree of coverage IP 67 according to ČSN EN 60529
Stem material stainless steel DIN 1.4301
Head material stainless steel DIN 1.4301
Working conditions ambient temperature: -30 to 100 ° C; -30 to 70 ° C with transducer relative humidity: max. 100% (at ambient temperature 25 ° C) atmospheric pressure: 70 to 107 kPa
Weight about 0.25 kg
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
These resistance temperature sensors are designed for contact temperature measurement of liquid or gaseous substances. The sensors are manufactured in two variants: a small head with a resistance output signal and a large head with an output signal of 4 to 20 mA, as well as in versions with a bushing or with a connector. These sensors are primarily intended for use in the food industry. In combination with the central holder, the sensor is suitable for measuring the temperature in air conditioning ducts. The combination of sensor and well is suitable for measuring in pipes. The standard temperature range for the use of active sensors is -30 to 150 ° C, passive -30 to 200 ° C. The sensors can be used for all control systems that are compatible with the types of sensors or output signals listed in the table of technical parameters. Temperature sensors are designed for operation in a non-aggressive environment. ACCESSORIES DECLARATION, CERTIFICATION, CALIBRATION
  • The manufacturer issues an EU Declaration of Conformity.
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
  • Technical parameters
NS 170K NS 171K NS 172K NS 370K NS 371K 1 mA 1 mA 1 mA 0.3 mA 0.3 mA
Sensor type (K - with connector)
Sensor type Ni 1000/5000 Ni 1000/6180 Ni 891 Ni 10000/5000 Ni 10000/6180
Measuring range - 50 to 200 ° C (temperature around the head -30 to 100 ° C)
Max. dc measuring current
NS 173K PTS 170K PTS 270K PTS 370K HS 170K Pt 500/3850 0.7 mA 1.5 mA 1 mA
Sensor type (K - with connector)
Sensor type T1 = Ni 2226 Pt 100/3850 pT 1000/3850 NTC thermistor 20 kΩ
Measuring range -50 to 150 ° C - 50 to 200 ° C (temperature around the head -30 to 100 ° C) -30 to 150 ° C
Max. dc measuring current 3 mA 1 mW *)
*) maximum power consumption PTS 680K Pt 1000/3850 -td to 50 ° C -30 to 60 ° C 0 to 35 ° C 0 to 100 ° C 0 to 150 ° C
Sensor type (K - with connector) Note
Sensor type
Output signal 4 to 20 mA
Standard measuring ranges temperature around the head -30 to 70 ° C
Supply voltage (Unap) 11 to 30 V DC recommended value 24 V DC;
Load resistance Rz 150 Ω for Unap = 12 V 700 Ω for Unap = 24 V
Sensor interruption output signal > 24 mA
Sensor short circuit output signal <3.5 mA
OTHER PARAMETERS
Accuracy class Ni sensors: cl. B, Δt = ± (0.4 + 0.007t), for t ≥ 0; Δt = ± (0.4 + 0.028 | t |), for t ≤ 0 in ° C; Pt sensors: cl. B according to ČSN EN 60751, Δt = ± (0.3 + 0.005 | t |) in ° C NTC 20 kΩ: ± 1 ° C for the range 0 to 70 ° C
Measurement error <0.6% of range, minimum 0.5 ° C
Sensor connection according to the wiring diagram
Standard L1 stem length 70, 120, 180, 240 mm
Response time τ0.5 <9 s (in flowing water 0.4 m.s -1 )
Head connector type - sensors with connector Lumberg RSFM4, M 12
Insulation resistance > 200 MΩ at 500 V DC, 25 ° ± 3 ° C; humidity <85%
Degree of coverage IP 67 according to ČSN EN 60529
Stem material stainless steel DIN 1.4301
Head material stainless steel DIN 1.4301
Working conditions ambient temperature: -30 to 100 ° C; -30 to 70 ° C with transducer relative humidity: max. 100% (at ambient temperature 25 ° C) atmospheric pressure: 70 to 107 kPa
Weight about 0.25 kg
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
These resistance temperature sensors are designed for contact surface temperature measurement. The sensors are supplied with mounting tape and cap and are suitable for measuring the temperature on pipes. The plastic head is equipped with a cable gland (a terminal block is located in the head) or a connector. The standard temperature range of the sensors is -50 to 130 ° C. Thanks to the construction, the sensor itself is isolated from the influence of the surrounding environment. The sensors can be used for all control systems that are compatible with the types of sensors or output signals listed in the table of technical parameters. Easy installation of temperature sensors is ensured thanks to the unique " S head " design of SENSIT s.r.o. The sensors are designed for operation in a non-aggressive environment. ACCESSORIES DECLARATION, CERTIFICATION, CALIBRATION
  • The manufacturer issues an EU Declaration of Conformity.
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached files
NS 140 Ns 140K NS 141 NS 141K NS 142 NS 142K NS 340 NS 340K NS 341 NS 341K 1 mA 1 mA 1 mA 0.3 mA 0.3 mA
Sensor type (K - with connector)
Sensor type Ni 1000/5000 Ni 1000/6180 Ni 891 Ni 10000/5000 Ni 10000/6180
Measuring range - 50 to 130 ° C (ambient temperature -30 to 100 ° C)
Max. dc measuring current
NS 143 NS 143K PTS 240 PTS 240K PTS 340 PTS 340K HS 140 HS 140K Pt 500/3850 Pt 1000/3850 0.7 mA 1.5 mA 1 mA
Sensor type (K - with connector) PTS 140 PTS 140K
Sensor type T1 = Ni 2226 Pt 100/3850 NTC thermistor 20 kΩ
Measuring range - 50 to 130 ° C (temperature around the head -30 to 100 ° C)
Max. dc measuring current 3 mA 1 mW *)
*) maximum power consumption NS 540 NS 540K NS 740 NS 740K Pt 1000/3850 Pt 1000/3850 -td to 50 ° C -30 to 60 ° C 0 to 35 ° C 0 to 100 ° C 0 to 150 ° C -td to 60 ° C 0 to 35 ° C 0 to 100 ° C 0 to 150 ° C
Sensor type (K - with connector) Note
Sensor type
Output signal 4 to 20 mA 0 to 10 V
Standard measuring ranges **) temperature around the head -30 to 70 ° C; temperature around the sensor head NS 840 (K) -30 ° C to 70 ° C
Supply voltage (Unap) 11 to 30 V DC 15 to 30 V DC recommended value 24 V DC; recommended power supply 12 V DC NS 840 (K) Axima AXSP3P02012
Load resistance Rz 150 Ω for Unap = 12 V 700 Ω for Unap = 24 V > 10 kΩ
Sensor interruption output signal > 24 mA > 10.5 V
Short circuit output signalsensors <3.5 mA ~ 0 V
**) Based on the customer's request, it is possible to provide a custom measuring range in the range from -40 to 150 ° C, while the minimum range of the range must be 30 ° C (eg -20 to 15 ° C; -30 to 80 ° C, etc. ) 40 cm 20 mm
OTHER PARAMETERS
Sensor accuracy class Ni sensors: cl. B, Δt = ± (0.4 + 0.007t), for t ≥ 0; Δt = ± (0.4 + 0.028 | t |), for t ≤ 0 in ° C; Pt sensors: cl. B according to ČSN EN 60751, Δt = ± (0.3 + 0.005 | t |) in ° C NTC 20 kΩ: ± 1 ° C for the range 0 to 70 ° C
Measurement error NS 540 (K), NS 740 (K) and NS 840 (K) *), **) <0.6% of range, minimum 0.5 ° C NS 840 (K) 0.5 ° C for range <100 ° C, 100 ° C
Sensor connection according to the scheme
Response time τ 0.5 <13 s (on a smooth surface without paste)
Recommended wire cross-section - bushing sensors 0.35 to 1.5 mm 2
Head connector type - sensors with connector RSFM4 - Lumberg
Insulation resistance > 200 MΩ at 500 V DC, 25 ° ± 3 ° C; humidity <85%
Degree of coverage IP 65 according to ČSN EN 60529
Case Material brass
Head material POLYAMIDE
Protective case material POLYAMIDE
Standard tape length
Minimum pipe diameter
Working conditions ambient temperature: -50 to 100 ° C; -30 to 70 ° C with transducer; -30 to 70 ° C with freq. output relative humidity: max. 100% (at ambient temperature 25 ° C) atmospheric pressure: 70 to 107 kPa
Weight about 0.15 kg
*) The measurement error of the temperature sensor depends on the influence of the ambient temperature and the environment and the properties of the measured surface (method error). **) It is recommended to apply a heat-conducting paste or silicone grease to the measured surface, which ensures a faster response and minimizes the method error when measuring the temperature with an attached sensor. Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
These resistance temperature sensors are designed for contact surface temperature measurement from ferromagnetic materials. Contact with the measured surface is ensured by the springing of the measuring sleeve. The plastic head is equipped with a cable gland (a terminal block is located in the head) or a connector. The standard temperature range of the sensors is -30 to 100 ° C. Temperature sensors can be used for all control systems that are compatible with the types of sensors or output signals listed in the table of technical parameters. Temperature sensors are designed for operation in a non-aggressive environment. ACCESSORIES DECLARATION, CERTIFICATION, CALIBRATION
  • The manufacturer issues an EU Declaration of Conformity.
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
Possible alternatives of S 190 series sensors are cable sensors with magnetic attachment .
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached files
NS 190 NS 190K NS 191 NS 191K NS 192 NS 192K NS 390 NS 390K NS 391 NS 391K 1 mA 1 mA 1 mA 0.3 mA 0.3 mA
Sensor type (K - with connector)
Sensor type Ni 1000/5000 Ni 1000/6180 Ni 891 Ni 10000/5000 Ni 10000/6180
Measuring range - 30 to 100 ° C
Max. dc measuring current
NS 193 NS 193K PTS 190 PTS 190K PTS 290 PTS 290K HS 190 HS 190K Pt 500/3850 Pt 1000/3850 0.7 mA 1.5 mA 1 mA
Sensor type (K - with connector) PTS 390 PTS 390K
Sensor type T1 = Ni 2226 Pt 100/3850 NTC thermistor 20 kΩ
Measuring range - 30 to 100 ° C
Max. dc measuring current 3 mA 1 mW *)
*) maximum power consumption NS 590 NS 590K NS 790 NS 790K Pt 1000/3850 Pt 1000/3850 -td to 60 ° C 0 to 35 ° C 0 to 100 ° C 0 to 150 ° C -td to 60 ° C 0 to 35 ° C 0 to 100 ° C 0 to 150 ° C
Sensor type (K - with connector) Note
Sensor type
Output signal 4 to 20 mA 0 to 10 V
Standard measuring ranges **) temperature around the head -30 to 70 ° C
Supply voltage (Unap) 11 to 30 V DC 15 to 30 V DC recommended value 24 V DC
Load resistance Rz 150 Ω for Unap = 12 V 700 Ω for Unap = 24 V > 10 kΩ
Sensor interruption output signal > 24 mA > 10.5 V
Sensor short circuit output signal <3.5 mA ~ 0 V
**) Based on the customer's request, it is possible to provide a custom measuring range in the range from -40 to 150 ° C, while the minimum range of the range must be 30 ° C (eg -20 to 15 ° C; -30 to 80 ° Cetc.) 19 mm 5 N 60 N
OTHER PARAMETERS
Sensor accuracy class Ni sensors: cl. B, Δt = ± (0.4 + 0.007t), for t ≥ 0; Δt = ± (0.4 + 0.028 | t |), for t ≤ 0 in ° C; Pt sensors: cl. B according to ČSN EN 60751, Δt = ± (0.3 + 0.005 | t |) in ° C NTC 20 kΩ: ± 1 ° C for the range 0 to 70 ° C
Measurement error NS 590 (K), NS 790 (K) <0.6% of range, minimum 0.5 ° C
Sensor connection according to the wiring diagram
Diameter of the measuring sleeve
Measuring sleeve pressure
Sensor holding force
Recommended wire cross-section - bushing sensors 0.35 to 1.5 mm 2
Head connector type - sensors with connector RSFM4 - Lumberg
Insulation resistance > 200 MΩ at 500 V DC, 25 ° ± 3 ° C; humidity <85%
Degree of coverage IP 52 according to ČSN EN 60529
Head material LEXAN 500R
Working conditions ambient temperature: -30 to 100 ° C; -30 to 70 ° C with transducer relative humidity: max. 100% (at ambient temperature 25 ° C) atmospheric pressure: 70 to 107 kPa
Weight about 0.1 kg (NS 590, NS 790); 0.08kg (passive)
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
These resistance temperature sensors are designed for contact surface temperature measurement. They are supplied with fastening tape and cap and are suitable for measuring the temperature on pipes. The dimensions of the sensor also allow placement under the pipe insulation. The standard temperature range of the sensors is -50 to 130 ° C. Thanks to the construction, the sensor itself is isolated from the influence of the surrounding environment. The sensors can be used for all control systems that are compatible with the types of sensors or output signals listed in the table of technical parameters. The sensors are designed for operation in a non-aggressive environment ACCESSORIES DECLARATION, CERTIFICATION, CALIBRATION
  • The manufacturer issues an EU Declaration of Conformity.
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached files
NS 150A NS 151A NS 152A NS 350A NS 351A 1 mA 1 mA 1 mA 0.3 mA 0.3 mA
Sensor type
Sensor type Ni 1000/5000 Ni 1000/6180 Ni 891 Ni 10000/5000 Ni 10000/6180
Measuring range - 50 to 130 ° C
Max. dc measuring current
NS 153A PTS 150A PTS 350A HS 150A Pt 500/3850 Pt 1000/3850 0.7 mA 1.5 mA 1 mA
Sensor type PTS 250A
Sensor type T1 = Ni 2226 Pt 100/3850 NTC thermistor 20 kΩ
Measuring range - 50 to 130 ° C
Max. dc measuring current 3 mA 10 mW *)
*) maximum power consumption 2.5, 10 m 40 cm 20 mm
OTHER PARAMETERS
Sensor accuracy class **) Ni sensors: cl. B, Δt = ± (0.4 + 0.007t), for t ≥ 0; Δt = ± (0.4 + 0.028 | t |), for t ≤ 0 in ° C Pt sensors: cl. B according to ČSN EN 60751, Δt = ± (0.3 + 0.005 | t |) in ° C NTC 20 kΩ: ± 1 ° C for the range 0 to 70 ° C
Sensor connection according to the wiring diagram
Response time variant S150A τ0.5 ≤ 10 s, τ0.9 ≤ 45 s (on a flat surface of Al prism without paste)
Insulation resistance > 200 MΩ at Vss, 25 ° ± 3 ° C; humidity <85%
Degree of coverage IP 65 according to ČSN EN 60529
Case Material brass
Power cord silicone shielded 2 x 0.22 mm 2 shielded silicone 4 x 0.15 mm 2
Standard cable length
Protective case material POLYAMIDE variant S150A
Standard tape length
Minimum pipe diameter
Working conditions ambient temperature: -50 to 130 ° C relative humidity: max. 85% (at ambient temperature 25 ° C) atmospheric pressure: 87 to 107 kPa
Weight about 0.15 kg
**) Checked by immersing the sensors in the liquid, no method error is considered - ambient influence temperatures and environments, surface irregularities, etc. must be added for two-wire connection also the influence of the resistance of the supply cable line, at a temperature of 20 ° C the influence of the resistance of the conductors is 0.4 ° C / 1 m. Note: Subject to technical changes bez alert. Due to technical developments during the year, there may be small changes in the products sold and their properties. For up-to-date information, please always contact the dealer.

Temperature sensors with cable

Description and usage
These resistance temperature sensors , cable temperature sensors, are designed for contact temperature measurement of gaseous, liquid or solid substances. The maximum temperature range of use is -50 ° C to 200 ° C. The 1.5 mm housing diameter ensures fast response to temperature changes and temperature measurements in hard-to-reach places. Wires with Teflon insulation without shielding are used as supply conductors and their length, including the housing, is 100 mm. The supply wires can be extended with different types of cables. Temperature sensors with cable are designed for universal use in chemically non-aggressive environments. The method of use must be chosen with regard to the temperature and chemical resistance of the housing and the supply cable. ACCESSORIES - connectors - extension cables DECLARATION, CERTIFICATION, CALIBRATION
  • EU Declaration of Conformity - according to Act 22/1997 Coll. on technical requirements for products, as amended.
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached files
TG 11.5 Pt 100, Pt 1000 1.5 mm
Sensor
Temperature range of use -50 ° C to 200 ° C
Sensor type
Sensor connection 2-wire
Connect extension cables 2-wire, 3-wire, 4-wire
Degree of coverage IP 67 according to ČSN EN 60 529
Case Material stainless steel DIN 1.4301
Case diameter
Case length L 20 to 40 mm
Lead wires 2 x AWG 32
Wire length 80 mm (for housing length 20 mm)
Response time 0.5 = 1.5 s; τ 0.9 = 4.5 s
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
These resistance temperature sensors, cable temperature sensors, are designed for contact temperature measurement of gaseous, liquid or solid substances. The temperature range of use is -50 to 200 ° C. The housing diameter of 2 mm ensures a fast response to temperature changes and temperature measurements in hard-to-reach places. The type with Teflon insulation without shielding is used as the supply cable. Temperature sensors with cable are designed for universal use, the method of use must be chosen with regard to the temperature and chemical resistance of the housing and the supply cable. Cable temperature sensors are designed for operation in a chemically non-aggressive environment. ACCESSORIES DECLARATION, CERTIFICATION, CALIBRATION
  • EU Declaration of Conformity - according to Act 22/1997 Coll. on technical requirements for products, as amended.
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached files
TG 12 Pt 100, Pt 500, Pt 1000
Sensor
Temperature range of use -50 ° C to 200 ° C
Sensor type
Degree of coverage IP 67 according to ČSN EN 60 529
Case Material stainless steel DIN 1.4301
Case diameter 2 mm
Case length L 25 to 60 mm
Power cord Teflon unshielded 4 x 0.02 mm 2 (4 x AWG 34)
Line resistance 0.765 Ω / m (resistance of one 1 m long wire)
Response time 0.5 = 1.5 s; τ 0.9 = 4.5 s
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
These resistance sensors are designed for contact temperature measurement of gaseous, liquid or solid substances. The temperature range of use for variant TG 3 is -50 ° C to 200 ° C and for variant TG 3A it is -50 ° C to 260 ° C. The housing diameter of 3 mm ensures a fast response to temperature changes, the type with Teflon insulation without shielding is used as the supply cable. The sensors are designed for universal use, the method of use must be chosen with regard to the temperature and chemical resistance of the housing and the supply cable. The sensors are designed for operation in a chemically non-aggressive environment. ACCESSORIES
  • Screw connection with collet or cutting rings - if different immersion lengths of the temperature sensor are set
  • connectors
DECLARATION, CERTIFICATION, CALIBRATION
  • EU Declaration of Conformity - according to Act 22/1997 Coll. on technical requirements for products, as amended.
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached Files
TG 3 TG 3A -td ° -50 ° C (short-term up to 300 ° C)
Sensor
Temperature range of use -50 ° C to 200 ° C
Sensor type PT, Ni, NTC, Thermocouple K, J
Degree of coverage IP 67 according to ČSN EN 60 529 IP 64 according to ČSN EN 60 529
Case material stainless steel DIN 1.4301
Case diameter 3 mm
Case length L 25 to 60 mm
Power cord unshielded Teflon 2 x 0.205 mm 2 teflon shielded 4 x 0.051 mm 2
Line resistance 0.16 Ω per 1 m cable for 2-wire connection
Response time τ 0,5 = up to 3 s, τ 0,9 = up to 9 s
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.

With smooth metal housing

Sheathed resistance sensors

Description and usage
Sheath resistance temperature sensors ( MGO temperature sensors ) are designed to measure the temperature of gaseous, liquid or solid substances. Their main advantage over the standard design of resistance temperature sensors is the formability of the sensor body. The temperature range of the measuring end is up to 500 ° C depending on the design of the temperature sensor. These jacketed resistance temperature sensors ( MGO temperature sensors ) can be used for all control systems that are compatible with the sensor type listed in the specifications table. Sheathed resistance temperature sensors ( MGO temperature sensors ) comply with the degree of protection up to IP 67 according to ČSN EN 60 529 depending on the type of supply cable and the used production technology. The housing diameter of 1.5 mm ensures a very fast time response of the sensor to a change in temperature. Sheath resistance temperature sensors ( MGO temperature sensors ) are designed for operation in a chemically non-aggressive environment, the method of use must be chosen with regard to the temperature and chemical resistance of the housing and the supply cable. ACCESSORIES
  • connectors
  • Screw connection with collet or cutting rings - if different immersion lengths of the temperature sensor are set
DECLARATION, CERTIFICATION, CALIBRATION
  • The manufacturer issues an EU Declaration of Conformity .
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached files
Pt 100, Pt 500, Pt 1000 1.5 mm
Sensor type TR 091.5P
Sensor type
Degree of Coverage to IP67 according to the supply cable variant
Execution Sheath resistance sensor
Diameter of the measuring part
Length of the measuring part from 50 to 300 mm as standard, others as required
Maximum measuring range (relative to the measuring tip) -50 ° C to 500 ° C (may be limited according to the length of the measuring tip and the variant of the supply cable)
Accuracy class Class B standard (Class A on request)
Casing material stainless steel DIN1.4571
Sensor body flexible, malleable
Minimum bend radius 10 times the average
Power cord variants / cable temperature resistance PVC shielded -30 ° C to 80 ° C Unshielded PVC -40 ° C to 105 ° C Shielded silicone -50 ° C to 200 ° C Teflon shielded -50 ° C to 250 ° C With glass insulation 0 ° C to 400 ° C (with metal braid)
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
Sheath resistance temperature sensors ( MGO temperature sensors ) are designed to measure the temperature of gaseous, liquid or solid substances. Their main advantage over the standard design of resistance temperature sensors is the formability of the sensor body. The temperature range of the measuring end is up to 500 ° C depending on the design of the temperature sensor. These jacketed resistance temperature sensors ( MGO temperature sensors ) can be used for all control systems that are compatible with the sensor type listed in the specifications table. Sheathed resistance temperature sensors ( MGO temperature sensors ) comply with the degree of protection up to IP 67 according to ČSN EN 60 529 depending on the type of supply cable and the used production technology. The housing diameter of 2 mm ensures a very fast time response of the sensor to a change in temperature. Sheath resistance temperature sensors ( MGO temperature sensors ) are designed for operation in a chemically non-aggressive environment, the method of use must be chosen with regard to the temperature and chemical resistance of the housing and the supply cable. ACCESSORIES
  • connectors
  • Screw connection with collet or cutting rings - if different immersion lengths of the temperature sensor are set
DECLARATION, CERTIFICATION, CALIBRATION
  • The manufacturer issues an EU Declaration of Conformity .
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached files
Pt 100, Pt 500, Pt 1000
Sensor type TR 092.0P
Sensor type
Degree of Coverage to IP67 according to the supply cable variant
Execution Sheath resistance sensor
Diameter of the measuring part 2 mm
Length of the measuring part from 50 to 300 mm as standard, others as required
Maximum measuring range (relative to the measuring tip) -50 ° C to 500 ° C (may be limited according to the length of the measuring tip and the variant of the supply cable)
Accuracy class Class B standard (Class A on request)
Casing material stainless steel DIN1.4571
Sensor body flexible, malleable
Minimum bend radius 10 times the average
Power cord variants / cable temperature resistance PVC shielded -30 ° C to 80 ° C Unshielded PVC -40 ° C to 105 ° C Shielded silicone -50 ° C to 200 ° C Teflon shielded -50 ° C to 250 ° C With glass insulation 0 ° C to 400 ° C (with metal braid)
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
Sheath resistance temperature sensors ( MGO temperature sensors ) are designed to measure the temperature of gaseous, liquid or solid substances. Their main advantage over the standard design of resistance temperature sensors is the formability of the sensor body. The temperature range of the measuring end of the sensor is up to 600 ° C depending on the design of the temperature sensor. These jacketed resistance temperature sensors ( MGO temperature sensors ) can be used for all control systems that are compatible with the sensor type listed in the specifications table. Sheathed resistance temperature sensors ( MGO temperature sensors ) comply with the degree of protection up to IP 67 according to ČSN EN 60 529 depending on the type of supply cable and the used production technology. The housing diameter of 3 mm ensures a very fast time response of the sensor to a change in temperature. Sheath resistance temperature sensors ( MGO temperature sensors ) are designed for operation in a chemically non-aggressive environment, the method of use must be chosen with regard to the temperature and chemical resistance of the housing and the supply cable. ACCESSORIES
  • connectors
  • Screw connection with collet or cutting rings - if different immersion lengths of the temperature sensor are set
DECLARATION, CERTIFICATION, CALIBRATION
  • The manufacturer issues an EU Declaration of Conformity .
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached files
Pt 100, Pt 500, Pt 1000
Sensor type TR 093.0P
Sensor type
Degree of Coverage to IP67 according to the supply cable variant
Execution Sheath resistance sensor
Diameter of the measuring part 3 mm
Length of the measuring part from 20 to 300 mm as standard, others as required
Measuring range (relative to measuring tip) -50 to 600 ° C (may be limited according to the length of the measuring tip and the variant of the supply cable)
Accuracy class Class B standard (Class A on request)
Casing material stainless steel DIN1.4571
Sensor body flexible, malleable
Minimum bend radius 10 times the average
Power cord variants / cable temperature resistance PVC shielded -30 ° C to 80 ° C Unshielded PVC -40 ° C to 105 ° C Shielded silicone -50 ° C to 200 ° C Teflon shielded -50 ° C to 250 ° C With glass insulation 0 ° C to 400 ° C (with metal braid)
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
Sheath resistance temperature sensors ( MGO temperature sensors ) are designed to measure the temperature of gaseous, liquid or solid substances. Their main advantage over the standard design of resistance temperature sensors is the formability of the sensor body. The temperature range of the measuring end of the sensor is up to 600 ° C depending on the design of the temperature sensor. These jacketed resistance temperature sensors ( MGO temperature sensors ) can be used for all control systems that are compatible with the sensor type listed in the specifications table. Sheathed resistance temperature sensors ( MGO temperature sensors ) comply with the degree of protection up to IP 67 according to ČSN EN 60 529 depending on the type of supply cable and the used production technology. The housing diameter of 4 mm ensures a very fast time response of the sensor to a change in temperature. Sheath resistance temperature sensors ( MGO temperature sensors ) are designed for operation in a chemically non-aggressive environment, the method of use must be chosen with regard to the temperature and chemical resistance of the housing and the supply cable. ACCESSORIES
  • connectors
  • Screw connection with collet or cutting rings - if different immersion lengths of the temperature sensor are set
DECLARATION, CERTIFICATION, CALIBRATION
  • The manufacturer issues an EU Declaration of Conformity .
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached files
Pt 100, Pt 500, Pt 1000 4 mm
Sensor type TR 094.5P
Sensor type
Degree of Coverage to IP67 according to the supply cable variant
Execution Sheath resistance sensor
Diameter of the measuring part
Length of the measuring part from 20 to 300 mm as standard, others as required
Maximum measuring range (relative to the measuring tip) -50 to 600 ° C (may be limited according to the length of the measuring tip and the variant of the supply cable)
Accuracy class Class B standard (Class A on request)
Casing material stainless steel DIN1.4571
Sensor body flexible, malleable
Minimum bend radius 10 times the average
Power cord variants / cable temperature resistance PVC shielded -30 ° C to 80 ° C Unshielded PVC -40 ° C to 105 ° C Shielded silicone -50 ° C to 200 ° C Teflon shielded -50 ° C to 250 ° C With glass insulation 0 ° C to 400 ° C (with metal braid)
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
Sheath resistance temperature sensors ( MGO temperature sensors ) are designed to measure the temperature of gaseous, liquid or solid substances. Their main advantage over the standard design of resistance temperature sensors is the formability of the sensor body. The temperature range of the measuring end of the sensor is up to 600 ° C depending on the design of the temperature sensor. These jacketed resistance temperature sensors ( MGO temperature sensors ) can be used for all control systems that are compatible with the sensor type listed in the specifications table. Sheathed resistance temperature sensors ( MGO temperature sensors ) comply with the degree of protection up to IP 67 according to ČSN EN 60 529 depending on the type of supply cable and the used production technology. The diameter of the sensor is 6 mm. Sheath resistance temperature sensors ( MGO temperature sensors ) are designed for operation in a chemically non-aggressive environment, the method of use must be chosen with regard to the temperature and chemical resistance of the housing and the supply cable. ACCESSORIES
  • connectors
  • Screw connection with collet or cutting rings - if different immersion lengths of the temperature sensor are set
DECLARATION, CERTIFICATION, CALIBRATION
  • The manufacturer issues an EU Declaration of Conformity .
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached files
Pt 100, Pt 500, Pt 1000 6 mm
Sensor type TR 096.0P
Sensor type
Degree of Coverage to IP67 according to the supply cable variant
Execution Sheath resistance sensor
Diameter of the measuring part
Length of the measuring part from 20 to 300 mm as standard, others as required
Maximum measuring range (relative to the measuring tip) -50 to 600 ° C (may be limited according to the length of the measuring tip and the variant of the supply cable)
Accuracy class Class B standard (Class A on request)
Casing material stainless steel DIN1.4571
Sensor body flexible, malleable
Minimum bend radius 10 times the average
Power cord variants / cable temperature resistance PVC shielded -30 ° C to 80 ° C Unshielded PVC -40 ° C to 105 ° C Shielded silicone -50 ° C to 200 ° C Teflon shielded -50 ° C to 250 ° C With glass insulation 0 ° C to 400 ° C (with metal braid)
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.

Temperature sensors with bayonet

Description and usage
These resistance temperature sensors with bayonet TG 60, TG 61 and TG 62 are preferably designed for measuring the temperature of solids, but can also be used for measuring the temperature of liquids or gaseous substances. A bayonet nut is used to fasten the temperature sensor to the measuring point, which is screwed onto the spring. By combining a bayonet nut, a spring and a corresponding bayonet cap, it is possible to ensure a higher pressure of the temperature sensor with the bayonet in the measured place or to correct any changes in the distance between the nut and the measured place during the measurement. The maximum temperature range of the sensors is -50 ° C to 350 ° C (for a short time 400 ° C). The range is narrowed for the individual design variants by the type of temperature sensor and supply cable. Temperature sensors comply with the degree of protection from IP 50 to IP 67 according to ČSN EN 60 529, according to the variant of the supply cable. Temperature sensors with a bayonet are designed for operation in a chemically non-aggressive environment, the method of use must be chosen with regard to the temperature and chemical resistance of the housing and the supply cable. ACCESSORIES DECLARATION, CERTIFICATION, CALIBRATION
  • EU Declaration of Conformity - according to Act 22/1997 Coll. on technical requirements for products, as amended.
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
MODIFICATIONS AND CUSTOMIZATION The following parameters can be modified for standard bayonet temperature sensors:
  • Possibility to encapsulate two temperature sensors
  • Variant design of housing dimensions in the area: diameter - 5 to 7 mm, length L, housing material, housing end design
  • spring length
  • Accuracy class A (except Ni 10000/5000, Ni 10000/6180, T1 = Ni 2226 sensors, NTC 20 k term thermistor)
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached files
TG 60 TG 61 TG 62
Sensor type
Case termination straight hemisphere R6 tip 120 °
Sensor type all types
Case material stainless steel DIN 1.4301
Case length 10 to 85 mm (with 15 mm addition for cable fastening)
Case diameter 6 mm
Insulation resistance 200 MΩ at 500 V DC
Power cable variants / temperature range (may be limited by sensor type - specified in the manual) PVC shielded -30 ° C to 80 ° C Unshielded PVC -40 ° C to 105 ° C Shielded silicone -50 ° C to 200 ° C Teflon shielded -50 ° C to 250 ° C With glass insulation 0 ° C to 400 ° C (with metal braid)
Degree of coverage PVC and silicone cables IP 67 Teflon cable IP 64 Glass insulated cable IP 50
Material / Dimensions of the bayonet nut nickel-plated brass / L = 16mm, inner 12.8 mm
Material / Dimensions of the spring stainless steel DIN 1.4301 / L = 200mm, outer ∅ 6 mm, wire 0,7 0.7mm
Maximum allowable static cable pull 1 kg
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
These resistance temperature sensors TG 80, TG 81 and TG 82 are preferably designed for measuring the temperature of solids, but can also be used to measure the temperature of liquids or gaseous substances. A bayonet nut is used to fasten the sensor to the measuring point, which is screwed onto the spring. By combining a bayonet nut, a spring and a corresponding bayonet cap, it is possible to ensure a higher pressure of the temperature sensor with the bayonet in the measured place or to correct any changes in the distance between the nut and the measured place during the measurement. The maximum temperature range of the sensors is -50 ° C to 350 ° C (for a short time 400 ° C). The range is narrowed for the individual design variants by the type of temperature sensor and supply cable. Temperature sensors with bayonet meet the degree of protection from IP 50 to IP 67 according to ČSN EN 60 529, according to the variant of the supply cable. Temperature sensors with a bayonet are designed for operation in a chemically non-aggressive environment, the method of use must be chosen with regard to the temperature and chemical resistance of the housing and the supply cable. ACCESSORIES DECLARATION, CERTIFICATION, CALIBRATION
  • EU Declaration of Conformity - according to Act 22/1997 Coll. on technical requirements for products, as amended.
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
MODIFICATIONS AND CUSTOMIZATION The following parameters can be modified for standard bayonet temperature sensors:
  • Possibility to encapsulate two temperature sensors
  • Variant design of housing dimensions in the area: diameter - 5 to 7 mm, length L, housing material, housing end design
  • spring length
  • Accuracy class A (except Ni 10000/5000, Ni 10000/6180, T1 = Ni 2226 sensors, NTC 20 k term thermistor)
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached files
TG 80 TG 81 TG 82
Sensor type
Case termination straight hemisphere R8 tip 120 °
Sensor type Pt, Ni, NTC, TC, etc.
Insulation resistance 200 MΩ at 500 V DC
Case Material stainless steel DIN 1.4301
Case length 10 to 85 mm (with 15 mm addition for cable fastening)
Case diameter 8 mm
Power cord / temperature variants range (may be limited by type sensors - specified in the manual) PVC shielded -30 ° C to 80 ° C Unshielded PVC -40 ° C to 105 ° C Shielded silicone -50 ° C to 200 ° C Teflon shielded -50 ° C to 250 ° C With glass insulation 0 ° C to 400 ° C (with metal braid)
Degree of coverage PVC and silicone cables IP 67 Teflon cable IP 64 Glass insulated cable IP 50
Material / Dimensions of the bayonet nut Nickel-plated brass / L = 18mm, inner ∅15 mm
Spring Material / Dimensions stainless steel DIN 1.4301 / L = 200mm, outer ∅ 8 mm, Mm wire 0.9mm
Maximum allowable static cable pull 1 kg
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
These resistance temperature sensors with TG 55 bayonet are designed for measuring the temperature of solids, but they can also be used to measure the temperature of liquids or gaseous substances. A bayonet head, which is screwed onto the spring, is used to fasten the temperature sensor to the measuring point. By combining a bayonet nut, a spring and a corresponding bayonet adapter, it is possible to ensure a higher pressure of the temperature sensor with the bayonet to the measuring point or to correct any changes in the distance between the nut and the measuring point during the measurement. The maximum temperature range of the sensors is - 50 ° C to 350 ° C (for a short time 400 ° C). The range is narrowed for the individual design variants by the type of temperature sensor and supply cable. Temperature sensors comply with the degree of protection from IP 50 to IP 67 according to ČSN EN 60 529, according to the variant of the supply cable. Temperature sensors with bayonet are designed for operation in a chemically non-aggressive environment. ACCESSORIES
  • bayonet adapter
  • connectors
DECLARATION, CERTIFICATION, CALIBRATION
  • EU Declaration of Conformity - according to Act 22/1997 Coll. on technical requirements for products, as amended.
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached files
TG 55 6 mm 1 kg
Sensor type
Sensor type all types
Case length L 60 to 200 mm
Case diameter
Case termination straight, hemisphere R6 or tip 120 °
Case Material stainless steel DIN 1.4301
Rectangular material stainless steel
Power cord variants / temperature range (may be limited by sensor type - specified in the manual) PVC shielded -30 ° C to 80 ° C Unshielded PVC -40 ° C to 105 ° C Shielded silicone -50 ° C to 200 ° C Teflon shielded -50 ° C to 250 ° C With glass insulation 0 ° C to 400 ° C (with metal braid)
Degree of coverage IP 50 to IP 67 - depending on cable type
Bayonet Head Material / Dimensions nickel-plated brass / L = 16mm, inner 12.8 mm
Spring Material / Dimensions stainless steel DIN 1.4301 / L = 200mm, outer ∅ 8 mm, wire 1 1 mm
Insulation resistance > 200 MΩ at 500 VDC, 25 ° ± 3 ° C
Maximum allowable static cable pull
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.

Temperature probes

Description and usage
These resistance temperature probes S x091 / 250 are designed for measuring the temperature of bulk and solids with such a consistency that the metal tip is not damaged. They can also be used for measuring liquid or gaseous substances. The temperature range of use of the temperature probe is -50 ° C to 250 ° C (for a short time 300 ° C). The probes consist of a metal tip (stainless steel DIN 1.4301), a handle and a supply cable. The material composition of the probes complies with the requirements of Decree 38/2001 Coll., As amended, the construction allows easy cleaning and, as a result, temperature probes can be used in the food industry. The resistance sensor is located in the tip of the stainless steel housing. The probes are designed for operation in a chemically non-aggressive environment. DECLARATION, CERTIFICATION, CALIBRATION
  • EU Declaration of Conformity - according to Act 22/1997 Coll. on technical requirements for products, as amended.
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached files
4 mm 145 mm
Probe type S x091 / 250
Maximum temperature range of use -50 ° C to 250 ° C (short-term 300 ° C)
Sensor type All types
Degree of coverage IP 68 (1 bar) according to ČSN EN 60 529
Probe tip material stainless steel DIN1.4301
Probe tip diameter
Probe tip length
Handle material PEEK
Handle temperature resistance up to 250 ° C
Power cord Teflon shielded 2 x 0.14 mm 2 teflon shielded 4 x 0.14 mm 2
Line resistance 0.3 Ω per 1 m cable for 2-wire connection
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
These resistance temperature probes S x051 / 150 are designed for measuring the temperature of bulk and solids with such a consistency that the metal tip is not damaged. They can also be used for measuring liquid or gaseous substances. The temperature range of use of the temperature probe is -50 ° C to 150 ° C. The probes consist of a metal tip (stainless steel DIN 1.4301), a handle and a supply cable. The material composition of the probes complies with the requirements of Decree 38/2001 Coll., As amended, the construction allows easy cleaning and, as a result, temperature probes can be used in the food industry. The resistance sensor is located in the tip of the stainless steel housing. The probes are designed for operation in a chemically non-aggressive environment. DECLARATION, CERTIFICATION, CALIBRATION
  • EU Declaration of Conformity - according to Act 22/1997 Coll. on technical requirements for products, as amended.
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached files
4 mm 145 mm up to 150 ° C
Probe type S x051 / 150
Maximum temperature range of use -30 ° C to 150 ° C
Sensor type all types
Degree of coverage IP 68 (1 bar) according to ČSN EN 60529
Probe tip material stainless steel DIN 1.4301
Probe tip diameter
Probe tip length
Handle material silicone
Handle temperature resistance
Power cord silicone shielded 2 x 0.22 mm 2 shielded silicone 4 x 0.15 mm 2
Line resistance 0.16 Ω per 1 m cable for 2-wire connection
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
These S x061 / 200 resistance temperature probes are designed to measure the temperature of bulk and solids with such a consistency that the metal tip is not damaged. They can also be used for measuring liquid or gaseous substances. The temperature range of use of the temperature probe is -50 ° C to 200 ° C. The probes consist of a metal tip (stainless steel DIN 1.4301), a handle and a supply cable. The material composition of the probes complies with the requirements of Decree 38/2001 Coll., As amended, the construction allows easy cleaning and, as a result, temperature probes can be used in the food industry. The resistance sensor is located in the tip of the stainless steel housing. The probes are designed for operation in a chemically non-aggressive environment. DECLARATION, CERTIFICATION, CALIBRATION
  • EU Declaration of Conformity - according to Act 22/1997 Coll. on technical requirements for products, as amended.
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached files
125 mm up to 200 ° C
Probe type S x061 / 200
Maximum temperature range of use -50 ° C to 200 ° C
Sensor type All types
Degree of coverage IP 68 (1 bar) according to ČSN EN 60529
Probe tip material stainless steel DIN 1.4301
Probe tip diameter 4 mm (4.5 mm - two sensing elements)
Probe tip length
Handle material teflon
Handle temperature resistance
Power cord silicone shielded 2 x 0.22 mm 2 shielded silicone 4 x 0.15 mm 2
Line resistance 0.16 Ω per 1 m cable for 2-wire connection
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
These S x304 / 220 resistance temperature probes are designed to measure the temperature of bulk and solids with such a consistency that the metal tip is not damaged. They can also be used for measuring liquid or gaseous substances. The temperature range of use of the temperature probe is -50 ° C to 220 ° C. The probes consist of a metal tip (stainless steel DIN 1.4301), a handle and a supply cable. The material composition of the probes complies with the requirements of Decree 38/2001 Coll., As amended, the construction allows easy cleaning and, as a result, temperature probes can be used in the food industry. The resistance sensor is located in the tip of the stainless steel housing. The probes are designed for operation in a chemically non-aggressive environment. DECLARATION, CERTIFICATION, CALIBRATION
  • EU Declaration of Conformity - according to Act 22/1997 Coll. on technical requirements for products, as amended.
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached files
3.6 mm 120 mm
Probe type S x304 / 220
Maximum temperature range of use -50 ° C to 220 ° C
Sensor type All types
Degree of coverage IP 68 (1 bar) according to ČSN EN 60 529
Probe tip material stainless steel 17240 (equivalent to 1.4301)
Probe tip diameter
Probe tip length
Handle material teflon
Handle temperature resistance -50 ° C to 200 ° C
Power cord silicone shielded 2 x 0.22 mm 2
Line resistance 0.16 Ω per 1 m cable for 2-wire connection
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
These resistance temperature probes S x061 / 250 are designed for measuring the temperature of bulk and solids with such a consistency that the metal tip is not damaged. They can also be used for measuring liquid or gaseous substances. The temperature range of use of the temperature probe is -50 ° C to 250 ° C. The probes consist of a metal tip (stainless steel DIN 1.4301), a handle and a supply cable. The material composition of the probes complies with the requirements of Decree 38/2001 Coll., As amended, the construction allows easy cleaning and, as a result, temperature probes can be used in the food industry. The resistance sensor is located in the tip of the stainless steel housing. The probes are designed for operation in a chemically non-aggressive environment. DECLARATION, CERTIFICATION, CALIBRATION
  • EU Declaration of Conformity - according to Act 22/1997 Coll. on technical requirements for products, as amended.
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
  • Technical parameters
  • Dimensional sketch
  • Attached files
125 mm
Probe type S x061 / 250
Maximum temperature range of use -50 ° C to 250 ° C
Sensor type All types
Degree of coverage IP 68 (1 bar) according to ČSN EN 60 529
Probe tip material stainless steel DIN 1.4301
Probe tip diameter 4 mm (4.5 mm - two sensing elements)
Probe tip length
Handle material teflon
Handle temperature resistance up to 250 ° C
Power cord Teflon shielded 2 x 0.14 mm 2 teflon shielded 4 x 0.14 mm 2
Line resistance 0.3 Ω per 1 m cable for 2-wire connection
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.

Needle

Touch

Description and usage
These resistance temperature probes S x034 / 150 are designed for tactile measurement of solids with a flat and smooth surface. The temperature range of use of the temperature probe is -30 ° C to 150 ° C. The probes consist of a metal housing terminated with a special dish for contact with the measured surface, a handle and a supply cable. The resistance sensor is placed in a brass (duralumin) bowl, which is firmly placed in a special rubber housing, which reduces the influence of the environment on the measurement. The probes are designed for operation in a chemically non-aggressive environment. DECLARATION, CERTIFICATION, CALIBRATION
  • EU Declaration of Conformity - according to Act 22/1997 Coll. on technical requirements for products, as amended.
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached files
70 mm
Probe type S x034 / 150
Maximum temperature range of use -30 ° C to 150 ° C
Sensor type all types
Degree of coverage IP 43 according to ČSN EN 60529
Probe tip material stainless steel DIN 1.4301
Touchpad material brass
Probe housing length
Handle material teflon
Handle temperature resistance -30 ° C to 200 ° C
Power cord silicone shielded 2 x 0.22 mm 2
Line resistance 0.16 Ω per 1 m cable for 2-wire connection
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
These resistance temperature probes S x033 / 250 are designed for tactile measurement of solids with a flat and smooth surface. The temperature range of use of the temperature probe is -30 ° C to 250 ° C. The probes consist of a metal housing terminated with a special dish for contact with the measured surface, a handle and a supply cable. The resistance sensor is placed in a brass (duralumin) bowl, which is firmly placed in a special rubber housing, which reduces the influence of the environment on the measurement. The probes are designed for operation in a chemically non-aggressive environment. DECLARATION, CERTIFICATION, CALIBRATION
  • EU Declaration of Conformity - according to Act 22/1997 Coll. on technical requirements for products, as amended.
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached files
125 mm up to 200 ° C
Probe type S x033 / 250
Maximum temperature range of use -30 ° C to 250 ° C
Sensor type All types
Degree of coverage IP 43 according to ČSN EN 60 529
Probe tip material stainless steel 17240 (equivalent to 1.4301)
Touchpad material dural 424254.61
Probe housing length
Handle material teflon
Handle temperature resistance
Power cord silicone shielded 2 x 0.22 mm 2 shielded silicone 4 x 0.15 mm 2
Line resistance 0.16 Ω per 1 m cable for 2-wire connection
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
These resistance temperature probes S x042 / 150 are designed for contact measurement of gaseous substances. The temperature range of use of the temperature probe is 0 ° C to 150 ° C. The probe consists of a metal housing with holes (food grade stainless steel 17240 corresponds to DIN 1.4301), handle, supply cable. The material composition of the probes complies with the requirements of Decree 38/2001 Coll., As amended, the construction allows easy cleaning and, as a result, temperature probes can be used in the food industry. The resistance sensor is located in the part with holes, which ensures direct contact with the measured gaseous substance and a fast response to temperature changes. The probes are designed for operation in a chemically non-aggressive environment. DECLARATION, CERTIFICATION, CALIBRATION
  • EU Declaration of Conformity - according to Act 22/1997 Coll. on technical requirements for products, as amended.
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached files
0 ° C to 150 ° C 120 mm
Probe type S x042 / 150
Maximum temperature range of use
Sensor type All types
Degree of coverage IP 20 according to ČSN EN 60 529
Probe tip material stainless steel 17240 (equivalent to 1.4301)
Probe housing diameter 5 mm
Probe housing length
Handle material teflon
Handle temperature resistance up to 250 ° C
Power cord silicone shielded 2 x 0.22 mm 2
Line resistance 0.16 Ω per 1 m cable for 2-wire connection
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
These temperature probes are designed for contact measurement of gaseous substances. The S x042 probe consists of a metal housing with holes, a handle, a standard 1 m supply cable and can be used for temperature measurement in the food industry due to the materials used. The resistance sensor is located in the part with holes, which ensures direct contact with the measured gaseous substance and a fast response to temperature changes. The design of the individual probes is shown in the figures. DECLARATION, CERTIFICATION, CALIBRATION
  • EU Declaration of Conformity - according to Act 22/1997 Coll. on technical requirements for products, as amended.
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
  • Technical parameters
  • Dimensional sketch
  • Attached files
Pttd - 3 mA; Pt 500 - 1.5 mA; Pt 1000, Ni 1000, Ni 891 - 1 mA; NTC 20kΩ - max. Power consumption 1 mW
Accuracy class Ni sensors: cl. B, Δt = ± (0.4 + 0.007t), for t≥ 0; Δt = ± (0.4 + 0.028 | t |), pro Pt sensors: cl. B according to ČSN EN 60 751, Δt = ± (0.3 + 0.005 | t |) in ° C NTC20kΩ: ± 1 ° C for 0 to 70 ° C range
Maximum DC measuring current
Involving probes 2-wire
Standard cable lengths 1, 2, 5, 10 m
200-80 / E ElKA K1321 200-80 / C CINCH
Sensor type Ni 1000/5000, Ni 1000/6180, Pt 100, Pt 1000
Tube material nickel-plated copper
Case Material S x042 stainless steel 17240
Handle material S x042 teflon
Connector type
Power cord silicone shielded 2 x 0.22 mm 2
Description and usage
These temperature probes are designed for contact measurement of gaseous substances. The resistance sensor is located in the part with holes, which ensures direct contact with the measured gaseous substance and a fast response to temperature changes. The design of the individual probes is shown in the figures. DECLARATION, CERTIFICATION, CALIBRATION
  • EC Declaration of Conformity - according to Act 22/1997 Coll. on technical requirements for products, as amended.
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
  • Technical parameters
  • Dimensional sketch
  • Attached files
Pttd - 3 mA; Pt 500 - 1.5 mA; Pt 1000, Ni 1000, Ni 891 - 1 mA; NTC 20kΩ - max. Power consumption 1 mW
Accuracy class Ni sensors: cl. B, Δt = ± (0.4 + 0.007t), for t≥ 0; Δt = ± (0.4 + 0.028 | t |), pro Pt sensors: cl. B according to ČSN EN 60 751, Δt = ± (0.3 + 0.005 | t |) in ° C NTC20kΩ: ± 1 ° C for 0 to 70 ° C range
Maximum DC measuring current
Involving probes 2-wire
Standard cable lengths 1, 2, 5, 10 m
200-80 / E ElKA K1321 200-80 / C CINCH
Sensor type Ni 1000/5000, Ni 1000/6180, Pt 100, Pt 1000
Tube material nickel-plated copper
Case Material S x042 stainless steel 17240
Handle material S x042 teflon
Connector type
Power cord silicone shielded 2 x 0.22 mm 2

Spatial


Accessories for thermometers

Sinks

Description and usage
JS 130 heat sinks are components that are usually screwed or welded into pipes. They are designed to protect the measuring part of the sensors against the effects of the measured environment. They enable the replacement of sensors during the operation of technological equipment. All types of sumps manufactured by SENSIT s.r.o. they are made of high quality stainless steel 17240 (corresponds to 1.4301). As standard, they are provided with a G 1/2 ″ thread, but it is possible to supply them with a thread M 20 x 1.5, M 27 x 2, G 3/8 ″, or others as required. JS 130 wells are designed primarily for temperature sensors with stem and plastic series S 120 (immersion sensors), temperature sensors series MINI , temperature sensors with stem and metal head (immersion sensors), all-stainless steel sensors heat PTS 580P, PTS580K . In addition to the standard lengths listed in the catalog sheets and in the table of technical parameters below, the sumps are also produced in non-standard lengths. The temperature sensors are locked in the well with an M4 x 10 screw. The sumps are manufactured to a nominal pressure of PN63.
  • Technical parameters
  • Dimensional sketch
  • Attached files
MAXIMUM FLOW RATE OF THE MEASURED MEDIA - AIR AND WATER STEAM / WATER [m.s -1 ] 15 / 1.5 8.0 / 1.0 2.5 / 0.6 0.6 / 0.3 35 / 3.5 30 / 3.0 15 / 2.0 5.0 / 1.2 1.6 / 0.8
Well diameter (mm) / well length (mm) to 60 > 60 to 100 > 100 to 160 > 160 to 220 > 220 to 400
∅6 to ∅ 8 20 / 2.0
∅ 10 to ∅ 12
JS 130 6.3 mm 10 mm
Well type
Well material stainless steel DIN 1.4301
Maximum sensor stem diameter
Well outer diameter
Standard L2 lengths 50, 100, 160, 220, 280, 340 and 400 mm
Standard threads G 1/2 ″; M 20 x 1.5; M 27 x 2; G 3/8 ″
Temperature range of well use
Designed for sensors temperature sensors with stem and plastic (immersion sensors), temperature sensors with stem and metal head (immersion sensors), all-stainless steel heat sensors PTS 580P, PTS580K
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
JS 130B heat sinks are components that are usually screwed or welded into pipes. They are designed to protect the measuring part of the sensors against the effects of the measured environment. They enable the replacement of sensors during the operation of technological equipment. All types of sumps manufactured by SENSIT s.r.o. they are made of high quality stainless steel 17240 (corresponds to 1.4301). As standard, they are provided with a G 1/2 ″ thread, but it is possible to supply them with a thread M 20 x 1.5, M 27 x 2, G 3/8 ″, or others as required. Wells JS 130B are designed primarily for for temperature sensors with stem and plastic head (immersion sensors), temperature sensors series MINI , temperature sensors with stem and metal head (immersion sensors), all-stainless steel sensors heat PTS 580P, PTS580K . In addition to the standard lengths listed in the catalog sheets and in the table of technical parameters below, the sumps are also produced in non-standard lengths. The temperature sensors are locked in the well with an M4 x 10 screw. The sumps are manufactured to a nominal pressure of PN63.
  • Technical parameters
  • Dimensional sketch
  • Attached files
MAXIMUM FLOW RATE OF THE MEASURED MEDIA - AIR AND WATER STEAM / WATER [m.s -1 ] 15 / 1.5 8.0 / 1.0 2.5 / 0.6 0.6 / 0.3 35 / 3.5 30 / 3.0 15 / 2.0 5.0 / 1.2 1.6 / 0.8
Well diameter (mm) / well length (mm) to 60 > 60 to 100 > 100 to 160 > 160 to 220 > 220 to 400
∅6 to ∅ 8 20 / 2.0
∅ 10 to ∅ 12
JS 130B 7.0 mm 10 mm
Well type
Well material stainless steel DIN 1.4301
Maximum sensor stem diameter
Well outer diameter
Standard L2 lengths 50, 100, 160, 220, 280 and 400 mm
Standard threads G 1/2 ″; M 20 x 1.5; M 27 x 2; G 3/8 ″
Temperature range of well use
Designed for sensors temperature sensors with stem and plastic (immersion sensors), temperature sensors with stem and metal head (immersion sensors), all-stainless steel heat sensors PTS 580P, PTS580K
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
JS 130G heat sinks are components that are usually screwed or welded into pipes. They are designed to protect the measuring part of the sensors against the effects of the measured environment. They enable the replacement of sensors during the operation of technological equipment. All types of sumps manufactured by SENSIT s.r.o. they are made of high quality stainless steel 17240 (corresponds to 1.4301). As standard, they are provided with a G 1/2 ″ thread, but it is possible to supply them with a thread M 20 x 1.5, M 27 x 2, G 3/8 ″, or others as required. JS 130 wells are designed for cable temperature sensors with bayonet TG 60, TG 61, TG62 . In addition to the standard lengths listed in the catalog sheets and in the table of technical parameters below, the sumps are also produced in non-standard lengths. The sumps are manufactured for a nominal pressure of PN63.
  • Technical parameters
  • Dimensional sketch
  • Attached files
MAXIMUM FLOW RATE OF THE MEASURED MEDIA - AIR AND WATER STEAM / WATER [m.s -1 ] 15 / 1.5 8.0 / 1.0 2.5 / 0.6 0.6 / 0.3 35 / 3.5 30 / 3.0 15 / 2.0 5.0 / 1.2 1.6 / 0.8
Well diameter (mm) / well length (mm) to 60 > 60 to 100 > 100 to 160 > 160 to 220 > 220 to 400
∅6 to ∅ 8 20 / 2.0
∅ 10 to ∅ 12
JS 130G 6.2 mm 10 mm TG 60, TG 61, TG 62
Well type
Well material stainless steel DIN 1.4301
Maximum sensor stem diameter
Well outer diameter
Standard L2 lengths 50, 100, 160, 220, 280 and 400 mm
Standard threads G 1/2 ″; M 20 x 1.5; M 27 x 2; G 3/8 ″
Temperature range of well use
Designed for sensors
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
The straight weld is intended for mounting wells in pipes or for mounting a temperature sensor for measurement directly in the medium. As standard, we produce weldments made of stainless steel DIN 1.4301 or iron DIN 1.0254. The most commonly used threads are G 1/2 ″ and M 20 x 1.5. Based on the customer's wishes, the dimensions of the weldment, its material and the threads used can be changed.
  • Technical parameters
  • Dimensional sketch
  • Attached files
28 mm
Type Straight fitting
Standard material stainless steel DIN 1.4301 or iron DIN 1.0254
Average
Standard length L 50 and 100 mm
Standard threads G 1/2 ″; M 20 x 1.5
Modifications based on the customer's request, we can change the dimensions of the weldment, material and fittings
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
The oblique weld is intended for mounting wells in pipes or for mounting a temperature sensor for measurement directly in the medium. As standard, we produce weldments made of stainless steel DIN 1.4301 or iron DIN 1.0254. The most commonly used threads are G 1/2 ″ and M 20 x 1.5. Based on the customer's wishes, the dimensions of the weldment, its material and the threads used can be changed.
  • Technical parameters
  • Dimensional sketch
  • Attached files
28 mm
Type Inclined weldment
Standard material stainless steel DIN 1.4301 or iron DIN 1.0254
Average
Standard length L 50 and 70 mm
Standard threads G 1/2 ″; M 20 x 1.5
Modifications based on the customer's request, we can change the dimensions of the weldment, material and fittings
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.

Coils

Reducers

Description and usage
The reducer is used to mount a well or temperature sensor for measurement directly in the medium. Thanks to the reduction, we will change the original thread in the application to the required thread. The standard material is iron 11120. The most commonly used internal threads are G 1/2 ″ and M 20 x 1.5 and external M 27 x 2. The threads can be different, according to the customer's wishes.
  • Technical parameters
  • Dimensional sketch
  • Attached files
16 mm 20 mm 11 mm
Type Reduction
Standard material iron 11120
Standard female threads / OK G 1/2 ″; M 20 x 1.5 / OK 30
Female thread length
Standard male thread M 27 x 2
Outer thread length
Inside diameter D
Modifications based on the customer's request, we can change the dimensions of the reduction, material and types of fittings
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
The plastic holder is used to mount sensors, for example, in ventilation ducts. The plastic holder is part of the package of temperature sensors with stem and last head S 120 series (NS 120, NS 121…, PTS 120, PTS 220…) . It is designed for a temperature sensor stem diameter of 6 mm.
  • Technical parameters
  • Dimensional sketch
  • Attached files
55 mm 6 mm 20 mm
Type Plastic holder
Maximum average
Inside diameter
Height
Material POLYAMIDE
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
The metal holder is used to attach the temperature sensors to, for example, a wall or a device wall. Metal holder is most often used for temperature sensors for outdoor environments series K 110 (NK 110, NK 111…, PTK 110, PTK 210…) . The holder is designed for temperature sensors with a stem diameter of 6 mm. The temperature sensors are locked in the K 110 holder with an M4 x 10 screw.
  • Technical parameters
  • Dimensional sketch
  • Attached files
<110 K 110
Type
Dimensions according to the dimensional sketch
Material stainless steel 17240 (equivalent to 1.4301)
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
Plastic holder for Flow sensors SNP" href="http://www.sensit.cz/shop/sensors-of-flow-snp/54/a26aa9fc.html"> sensors a flow switches SNP a SPP . The most common use is to attach a flow sensor or switch to ventilation ducts. The sensor and the flow switches are mounted in the holder by means of a bushing.
  • Technical parameters
  • Dimensional sketch
  • Attached files
Type Plastic holder - flow
Maximum average
Inside diameter
Height
Material POLYAMIDE
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
The metal holder is used to mount temperature sensors, for example in ventilation ducts. Metal holder is most often used for temperature sensors with stem and metal head series K 120 (NK 120, NK 121…, PTK 120, PTK 220…) , then u temperature sensors with stem and plastic head series S 120 , MINI series A u temperature sensors with all-stainless steel head . The holder is designed for temperature sensors with a stem diameter of 6 mm. The temperature sensors are locked in the K 120 holder with an M4 x 10 screw.
  • Technical parameters
  • Dimensional sketch
  • Attached files
48 mm 6.3 mm 15 mm
Type Metal holder K 120
Maximum average
Inside diameter
Height
Material stainless steel 17240 (equivalent to 1.4301)
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
Plastic holder designed for PTSV 120 temperature and relative humidity sensors . The most common use is to attach a temperature and relative humidity sensor to ventilation ducts. The temperature and relative humidity sensor is mounted in the holder by means of a bushing.
  • Technical parameters
  • Dimensional sketch
  • Attached files
55 mm
Type Plastic moisture holder
Maximum average
Inside diameter
Height
Material POLYAMIDE
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.

Brackets


Flood sensors

Conductivity

Description and usage
Flood sensors ( level sensors ) consist of metal sensing electrodes and a plastic head in which the evaluation electronics are located. The connection of the 24VDC or 24VAC supply voltage and at the same time the idle state at the output of the relay or transistor is indicated by a green LED diode. A fault condition (creation of a conductive connection between the electrodes) is signaled by a red LED inside the box and a change in the state of the output transistor or relay contact. The relay output is selected by placing a jumper on the PCB. All metal parts are made of stainless steel DIN 1.4301, the basic length of the sensing electrodes is 50 mm. The supply cables are connected to the terminal board via glands that are part of the plastic head. The basic material of the head is POLYAMIDE. Flood sensors ( level sensors ) work on the principle of different conductivity (conductivity) of air and water. Water becomes conductive to some extent due to dissolved minerals and organic substances. Under normal conditions (rainwater) it has a conductivity from 5 mS / m towards higher values. From this value, the sensitivity of the flood sensor itself (haldina sensors) is set with a margin (less than 1 mS / m). When the electrodes are connected by a conductive medium, a current (of the order of µA) passes through the circuit, which is detected by the evaluation electronics, which signals the status by a red LED, activating the output transistor (open collector) or closing the relay contacts. Flood sensors ( level sensors ) are designed to signal fault conditions related to water leakage in industrial plants, but also in rooms, offices, or production halls in the range from 0 up to 80 ° C. Flood sensors (level sensors) comply with the degree of protection IP 65 according to ČSN EN 60 529 and meet the conditions for operation in a normal, chemically non-aggressive environment, the method of use must be chosen with regard to temperature resistance of the head and chemical resistance of the housing and sensor head. Possible apps:
  • Switch off the pump when the required level is reached
  • Flooding of objects with groundwater, floods, sewers
  • Flooding the sumps, switching on the pump when the level is reached
  • Signaling a water leak from the washing machine, boiler, darling, etc.
  • Flooding of the room due to various equipment failures - damaged filling supply to the toilet, overflowing bathtub, washbasin
  • Condensate monitoring in the ventilation duct
DECLARATION, CERTIFICATION, CALIBRATION
  • EC Declaration of Conformity - according to Act 22/1997 Coll. on technical requirements for products, as amended.
  • Technical parameters
  • Attached files
approx. 15 mA approx. 35 mA 6 A 125 g
Sensor type SHV1
Supply voltage 15 to 30 VDC / VAC for relay output; recommended 24 VDC / VAC 12 to 30 VDC / VAC for open collector output; recommended 24 VDC / VAC
Temperature range of use around the head 0 to 80 ° C sensing electrodes up to a maximum of 100 ° C
Output signal relay output or open collector output terminal
Max. no-load collection
Max. load collection
Max. switching current
Switching voltage up to 24 VDC / VAC
Alarms LED red - alarm status LED green - device in operation, alarm status inactive
Switching sensitivity alarm function active at fluid conductivity from 1 m S / m -1
Insulation resistance between electrodes ≥ 500 VDC (between electrodes not connected to the PCB)
Head Cover IP65 according to ČSN EN 60 529
Head material POLYAMIDE
Electrode material stainless steel class DIN 1.4301
Dimensions head 90 x 63 x 34 mm including electrodes 141 x 63 x 34 mm
Weight
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For up-to-date information, please always contact the dealer.
Description and usage
Flood sensors ( level sensors ) consist of a plastic head in which the evaluation electronics are located and an external probe consisting of a silicone cable, at the end of which a plastic box with sensing sensors is connected. electrodes. The metal electrodes are made of stainless steel DIN 1.4301, the length of the sensing electrodes itself is 50 mm, the basic length of the connecting cable is 2 m. A fault condition (creation of a conductive connection between the electrodes) is signaled by a red LED inside the box and a change in the state of the output transistor or relay contact. The relay output is selected by placing a jumper on the PCB. The power supply cables and output contacts are connected to the terminal board via glands that are part of the plastic head. The basic material of the head is POLYAMIDE. Flood sensors ( level sensors ) work on the principle of different conductivity (conductivity) of air and water. Water becomes conductive to some extent due to dissolved minerals and organic substances. Under normal conditions (rainwater) it has a conductivity from 5 mS / m-1 towards higher values. From this value, the sensitivity of the sensor itself is set with reserve (less than 1 mS / m). A current (of the order of µA) flows between the immersed electrodes, which is detected by the evaluation electronics, which signals the status by a red LED, activating the output transistor (open collector) and closing or opening the relay (both according to the jumper on jumper). Flood sensors ( level sensors ) with an external probe are designed to signal fault conditions related to water leakage in industrial plants, but also in rooms, offices, or in production halls in the range from 0 to 80 ° C. Flood sensors (level sensors) comply with the degree of protection IP 65 according to ČSN EN 60 529 and meet the conditions for operation in a normal, chemically non-aggressive environment, the method of use must be chosen with regard to temperature resistance of the head and chemical resistance of the housing and sensor head. DECLARATION, CERTIFICATION, CALIBRATION
  • EC Declaration of Conformity - according to Act 22/1997 Coll. on technical requirements for products, as amended.
  • Technical parameters
  • Attached files
approx. 15 mA approx. 35 mA 6 A
Sensor type SHV2
Supply voltage 15 to 30 VDC / VAC for relay output; recommended 24 VDC / VAC 12 to 30 VDC / VAC for open collector output; recommended 24 VDC / VAC
Temperature range of use around the head 0 to 80 ° C sensing electrodes up to a maximum of 100 ° C
Output signal relay output or open collector output terminal
Max. no-load collection
Max. load collection
Max. switching current
Switching voltage up to 24 VDC / VAC
Alarms LED red - alarm status LED green - device in operation, alarm status inactive
Switching sensitivity alarm function active at fluid conductivity from 1 m S / m -1
Insulation resistance between electrodes ≥ 500 VDC (between electrodes not connected to the PCB)
Head Cover IP65 according to ČSN EN 60 529
External probe coverage IP67 according to ČSN EN 60 529
Head material POLYAMIDE
Electrode material stainless steel class DIN 1.4301
External probe material SILAMID
External probe cable length basic 2 m, custom up to 15 m
External probe cable type unshielded silicone 2 x 0.22 mm 2
Dimensions head 90 x 63 x 34 mm exground probe 61 x 24 x 25 mm
Weight 190 g with external probe 2 m
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer
Description and usage
Optical level sensors are used, for example, to measure, regulate and control production in the food and pharmaceutical industries. Furthermore, they can be found in applications for sensing the filling and emptying of tanks, sensing the maximum and minimum liquid level, to protect the pump against idling and more. Accessories Design in two types:
  • TYPE "A" - Suitable for oily liquids (such as coolants, oil, diesel, milk)
  • TYPE "B" - Suitable for foam liquids (example: beer, Kofola soft drinks, and other liquids)
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached files
Technical parameters
Execution With cable With connector
Nominal voltage Un 24 VDC
Supply voltage Uc 12 - 30 VDC
Supply current Ic <25 mA
Output current Iz ≤ 200 mA
Switch-on indication LED
Reverse polarity resistance YES
Switching range (% tip immersion) <cca 50%
Switching frequency 35 Hz
Operating temperature range T - 25 ° C to 100 ° C - process -25 ° C to 80 ° C - surroundings
Case / Optics stainless steel / crystal glass
Maximum working pressure 2 MPa
Degree of coverage IP 68 IP 67
Connection type PVC cable 3 x 0.25 mm 2 ; 2 m M12 connector
Design in two types: TYPE "A" - Suitable for oily liquids (such as coolants, oil, diesel, milk) TYPE "B" - Suitable for foam liquids (e.g. beer, Kofola soft drinks, and other liquids) POS 187 311 A POS 187 311 B POS 187 411 A POS 187 411 B POS 187 312 A POS 187 312 B POS 187 412 A POS 187 412 B POS 187 314 A POS 187 314 B POS 187 414 A POS 187 414 B
Order number
Output Cable version Connector version
TYPE "A" TYPE "B" TYPE "A" TYPE "B"
NPN Switching
Expandable
PNP Switch POS 187 313 A POS 187 313 B POS 187 413 A POS 187 413 B
Expandable
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Popis a použití
Optické snímače hladiny se používají například pro měření, regulaci a řízení výroby v potravinářském a farmaceutickém průmyslu. Dále je můžeme nalézt v aplikacích snímání naplnění a vyprázdnění nádrží, snímání maximální a minimální hladiny kapaliny, pro ochranu pumpy proti chodu na prázdno a další. Příslušenství Provedení ve dvou typech:
  • TYP „A“ – Vhodný pro olejnaté tekutiny (například chladící emulze, olej, nafta, mléko …)
  • TYP „B“ – Vhodný pro tekutiny s pěnou (například: pivo, nealkoholické nápoje typu Kofola a ostatní tekutiny)
  • Popis zboží
  • Technické parametry
  • Schéma zapojení
  • Rozměrový náčrt
  • Přiložené soubory
 DESCRIPTION AND APPLICATION Optical liquid level sensors are intended for measurement, regulation and control of production in foodstuff (e.g. beverages production) or pharmaceutical industry. They are used also in applications of sensing of a tank filling or emptying, min. and max. liquid level detection, a pump dry run protection etc. Variants of optical liquid level sensors:
  • Type ,,A“ – suitable for oil liquids (e.g. coling emulsions, oil, crude, hydraulic oil, milk, etc.)
  • Type ,,B“ – suitable for liquids with foam (e.g. beer,  ,,cola“ type drinks) and other liquids (e.g. water)

Optical


Transducers

Temperature-Voltage

Description and usage
The STU transducers are designed to convert the signal of resistance temperature sensors based on Ni 1000/6180, Pt 100/3850 and Pt 1000/3850 to a unified signal 0 to 10 V. The transducers can be used for all control systems that are compatible with the output signal 0 up to 10 V. The box can be equipped with a terminal for mounting on a DIN rail. The standard measuring ranges of the transmitter are given in the table of technical parameters, the temperature range of the transmitter resistance is -30 ° C to 80 ° C and must not be exceeded even for a short time. The sensors are designed for operation in a chemically non-aggressive environment. DECLARATION, CERTIFICATION, CALIBRATION
  • EU Declaration of Conformity - according to Act 22/1997 Coll. on technical requirements for products, as amended.
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached files
Pt 1000/3850 -td ° -30 ° C 0 ° C to 35 ° C 0 ° C to 100 ° C 0 ° C to 150 ° C 0 ° C to 250 ° C -td ° -30 ° C 0 ° C to 35 ° C 0 ° C to 100 ° C 0 ° C to 150 ° C 0 ° C to 200 ° C 0 ° C to 400 ° C -td ° -30 ° C 0 ° C to 35 ° C 0 ° C to 100 ° C 0 ° C to 150 ° C 0 ° C to 200 ° C 0 ° C to 400 ° C
Converter type STU Ni STU Pt 100 STU Pt 1000
Sensor type Ni 1000/6180 Pt 100/3850
Output signal 0 to 10V
Supply voltage (Unap) 15 to 30 V DC (recommended value 24 V DC)
Standard measuring ranges
Terminal block cover IP 65 according to ČSN EN 60 529
Ambient temperature - 30 ° C to 80 ° C
Measurement error <0.6% of range, minimum 0.5 ° C
Load resistance Rz > 50 kΩ
Power consumption <8 mA
Sensor open output voltage > 14 V
Sensor short-circuit output voltage ~ 0 V
Sensor connection according to the wiring diagram
Recommended wire cross section 0.35 to 1.5 mm 2
Box Material POLYAMIDE
Weight 0.15 kg
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
STI and STID type converters are designed to convert the signal of resistance temperature sensors based on Ni 1000/6180, Pt 100/3850 and Pt 1000/3850 to a unified 4 to 20 mA signal. The transducers can be used for all control systems that are compatible with a 4 to 20 mA output signal. The standard measuring ranges of the transmitter are given in the table of technical parameters, the temperature range of the transmitter resistance is -30 to 70 ° C and must not be exceeded even for a short time. The sensors are designed for operation in a chemically non-aggressive environment. In terms of the design of the converter box, we distinguish two basic variants: 1. STI Pt and STI Ni converters : plastic box made of POLYAMIDE material (identical to eg S 120 sensor head), the box is equipped with a bracket for wall mounting or a terminal for DIN rail mounting, the terminal box cover complies with IP 65 according to ČSN EN 60 529. 2. STID Pt and STID Ni converters : plastic box made of TARFLON - IRY 2200 material meeting the requirements of UL 94 V-0, designed for direct mounting in DIN rail cabinets. DECLARATION, CERTIFICATION, CALIBRATION
  • E U Declaration of Conformity - according to Act 22/1997 Coll. on technical requirements for products, as amended.
  • Calibration - All production undergoes output metrological inspection, which is performed by comparison with standards or working gauges. The traceability of standards and working gauges is ensured in the sense of §5 of Act No. 505/1990 on Metrology. The manufacturer offers the possibility to supply sensors calibrated in the laboratory of SENSIT s.r.o. (according to the requirements of the ČSN EN ISO / IEC 17025 standard) or in AKL.
  • Technical parameters
  • Wiring diagram
  • Dimensional sketch
  • Attached files
STI Pt 100 STID Pt 100 Pt 1000/3850 -td ° -30 ° C 0 ° C to 35 ° C 0 ° C to 100 ° C 0 ° C to 150 ° C 0 ° C to 250 ° C -td ° -30 ° C 0 ° C to 35 ° C 0 ° C to 100 ° C 0 ° C to 150 ° C 0 ° C to 200 ° C 0 ° C to 400 ° C -td ° -30 ° C 0 ° C to 35 ° C 0 ° C to 100 ° C 0 ° C to 150 ° C 0 ° C to 200 ° C 0 ° C to 400 ° C
Converter type STI Ni STID Ni STI Pt 1000 STID Pt 1000
Sensor type Ni 1000/6180 Pt 100/3850
Output signal 4 to 20 mA
Supply voltage (Unap) 11 to 30 V DC for STI; 11 to 30 V DC for STID (recommended value 24 V DC)
Maximum Unap ripple 0.5%
Standard measuring ranges
Terminal block cover IP 65 (for STI), IP 20 (for STID) according to ČSN EN 60 529
Ambient temperature - 30 ° C to 80 ° C
Measurement error <0.6% of range, minimum 0.5 ° C
Load resistance Rz 50 (Unap - 9) Ω
Output current when sensor is interrupted > 24 mA
Sensor short-circuit output current <3.5 mA
Sensor connection according to the wiring diagram
Recommended wire cross section STI: 0.35 to 1.5 mm 2 STID: 0.35 to 4 mm 2
Box Material STI: POLYAMIDE STID: TARFLON - IRY2200
Working conditions ambient temperature: -30 ° C to 70 ° C relative humidity: max. 85% (at ambient temperature 25 ° C) atmospheric pressure: 87 to 107 kPa
Weight 0.15 kg
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.

Temperature-current

Programmable

Description and usage
PP and SH1 series transducers are programmable transducers to the head and are designed for conversion of industrial signals from a resistance temperature sensor (SH1, SLM) or thermocouple to a unified industrial signal of 4 to 20 mA.
  • Technical parameters
  • Attached files
PP 200 PMA 300 Pt 100/3850 Pt 1000/3850 By 100/5000 and 6180 By 1000/5000 i6180 Pt 100/3850 Pt 1000/3850 By 100/5000 and 6180 By 1000/5000 and 6180 Pt 100/3850 Pt 1000/3850 By 100/5000 and 6180 By 1000/5000 and 6180 thermocouple J, L, K, T, S, B, E… 0.15% 0.15% 0.15%
Converter type PP 300
Input
Connect an input 2 or 3-wire 2-wire -
Output 4 to 20 mA, programmable range
Measurement range any, min. 10 ° C range any, min. 10 ° C range by thermocouple type
Linearization yes yes yes
Power 8 to 28 V after output loop 8 to 28 V after output loop 8 to 30 V after output loop
Accuracy
Operating temperature range -25 ° C to 80 ° C -25 ° C to 80 ° C -25 ° C to 80 ° C
*) errors in percent are related to the margin Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
Programmable measuring transducers are designed for conversion of industrial signals from a resistance temperature sensor (PP210, PSTID) or thermocouple (PSTID) to a unified industrial signal of 4 to 20 mA.
  • Technical parameters
  • Attached files
SUG2 Pt 100/3850 Pt 1000/3850 By 100/5000 and 6180 By 1000/5000 and 6180 Pt 100/3850 Pt 1000/3850 Pt 100/6180 By 1000/6180 thermocouple J, L, K, T, S, B, E…
Converter type PP210
Input
Connect an input 2 or 3-wire 2 or 3-wire
Output 4 to 20 mA 4 to 20 mA
Galvanically isolated output no yes
Measurement range any, min. 10 ° C range any, min. range in ° C according to sensor type
Linearization yes implemented programmatically
Power 8 to 28 V after output loop 9 to 30 V
Operating temperature range -20 ° C to 80 ° C -20 ° C to 80 ° C
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
The PPL100 converter is designed to convert the signal from a resistance thermometer in the temperature range -55 to 640 ° C to a number in ASCII format.
  • Technical parameters
  • Attached files
Pt 100/3850, Pt 1000/3850, Ni 100/5000 and 6180 Ni 1000/5000 and 6180 (other by agreement, eg NTC 20K) 0.2 ° C
Converter type PPL 100
Input
Output RS485 (-55 ° C to 640 ° C)
Linearization yes
Power 8 to 20 V
Accuracy
Operating temperature range -20 ° C to 80 ° C
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.
Description and usage
Programmable converter designed for conversion of industrial signals from a resistance temperature sensor Pt 100/3850 to a unified industrial signal 4 to 20 mA. Due to its construction, it is especially suitable for connection to cable sensors.
  • Technical parameters
  • Attached files
0.2%
Input Pt 100/3850
Connect an input 3-wire
Connector M12 4-pin
Output 4 to 20 mA
Measurement range programmable, max. -50 ° C to 800 ° C, min. range 50 ° C
Power 8.5 to 32 V DC
Accuracy
Dimensions diameter 14 mm, length 55 mm
Operating temperature range -40 ° C to 80 ° C (for plastic body)
Note: Specifications are subject to change without notice. Due to technical developments during the year, there may be small changes in the products sold and their properties. For current information, please always contact the dealer.